FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 VOLUMETRIC INFUSION PUMP

MDR report key: 1832686 · Received September 13, 2010

Report

Report Number
6000001-2007-01870
Event Type
Malfunction
Date Received
September 13, 2010
Date of Event
December 27, 2006
Report Date
December 27, 2006
Manufacturer
BAXTER HEALTHCARE PTE.LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).EVALUATION SUMMARY:EVALUATION WAS PERFORMED AND THE CONDITION WAS CONFIRMED. INSPECTION OF THE PUMP REVEALED DEPLETED MAIN BATTERIES CAUSED FAILURE CODE 570:320:844:0000. THE MAIN BATTERIES WERE REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA, (B)(4).(B)(4).

Description of Event or Problem · 1

DURING SERVICE BY BAXTER, FAILURE CODE 570:320:844:0000 WAS FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE.LTD. N/A NA

Patients

Seq Age Sex Outcome Treatment
1