FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1832660 · Received September 13, 2010

Report

Report Number
6000001-2007-01070
Event Type
Malfunction
Date Received
September 13, 2010
Date of Event
December 1, 2006
Report Date
December 1, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).EVALUATION SUMMARY:THE REPORTED CONDITION OF THE PUMP NOT HOLDING A CHARGE WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THIS WAS CAUSED BY FAILURE CODE 570:320:844:0000, WHICH WAS CAUSED BY DEPLETED BATTERIES, THEREFORE THE PUMP'S BATTERIES WERE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA, (B)(4).CONTINUED FROM H9:6000001-2/25/05-004-C6000001-12/13/05-019-C

Description of Event or Problem · 1

THE FACILITY REPORTED A PUMP WITH A BATTERY THAT WILL NOT HOLD A CHARGE. IT IS UNKNOWN WHEN THIS OCCURRED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA