COLLEAGUE VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-00806
- Event Type
- Malfunction
- Date Received
- September 13, 2010
- Date of Event
- December 14, 2006
- Report Date
- December 14, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-7/20/05-015-C
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS EVALUATED AT A BAXTER SERVICE CENTER. THE REPORTED CONDITION OF FAILURE CODE 703 WAS CONFIRMED. THE USER INTERFACE MODULE PRINTED CIRCUIT BOARD WAS FOUND TO BE FUNCTIONING OUTSIDE SPECIFICATION WHICH CAUSED THE FAILURE TO OCCUR. THE CIRCUIT BOARD WAS REPLACED AND THE DEVICE WAS RETURNED TO THE CUSTOMER FULLY OPERATIONAL. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA (B)(4).
THE DEVICE WAS RETURNED FROM CUSTOMER FOR SERVICE. DURING SERVICE BY BAXTER TECHNICIAN, THE DEVICE ALARMED WITH FAILURE CODE 703. CUSTOMER REPORTED THERE WERE NO PATIENT INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |