COLLEAGUE VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-02221
- Event Type
- Malfunction
- Date Received
- September 13, 2010
- Date of Event
- December 29, 2006
- Report Date
- December 29, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON (B)(4) 2007. EVALUATION SUMMARY: THE DEVICE EVALUATION WAS COMPLETED AND THE DEPLETED BATTERY CONDITION WAS CONFIRMED DUE TO POTENTIALLY DAMAGED BATTERIES. BAXTER SERVICE TECHNICIAN INSTALLED NEW BATTERIES AND TESTED THE DEVICE. A REVIEW OF THE COMPLAINT HISTORY REVEALED SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT AND THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA, MDQ-CAPA-132. 6000001-2/25/05-004-C, 6000001-12/13/05-019-C.
THE DEVICE WAS RECEIVED FOR SERVICE. DURING SERVICE TESTING DEPLETED BATTERIES WERE FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |