FDA Adverse Event Injury Summary report: N

PROXIMATE*ILS CURVED CIR STAPL

MDR report key: 1832555 · Received September 13, 2010

Report

Report Number
3005075853-2010-05229
Event Type
Injury
Date Received
September 13, 2010
Date of Event
August 18, 2010
Report Date
August 19, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE AFFILIATE. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING RESPONSE WAS RECEIVED ON 9/8/2010: (B)(6). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. IN ADDITION, THE STAPLE RETAINER WAS RECEIVED IN GOOD VISUAL CONDITIONS. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, AFTER THE FIRING, ONLY HALF CIRCLE TISSUE WAS CUT AND ANASTOMOSED. THEN THE SURGEON USED HAND COLOSTOMY TO FINISH THE PROCEDURE. NOW THE PATIENT IS FINE. THE PATIENT HAD RECTUM CANCER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE*ILS CURVED CIR STAPL STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 76 YR Other| R