PROXIMATE*ILS CURVED CIR STAPL
Report
- Report Number
- 3005075853-2010-05229
- Event Type
- Injury
- Date Received
- September 13, 2010
- Date of Event
- August 18, 2010
- Report Date
- August 19, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K983536
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION WAS NOT PROVIDED BY THE AFFILIATE. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING RESPONSE WAS RECEIVED ON 9/8/2010: (B)(6). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. IN ADDITION, THE STAPLE RETAINER WAS RECEIVED IN GOOD VISUAL CONDITIONS. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, AFTER THE FIRING, ONLY HALF CIRCLE TISSUE WAS CUT AND ANASTOMOSED. THEN THE SURGEON USED HAND COLOSTOMY TO FINISH THE PROCEDURE. NOW THE PATIENT IS FINE. THE PATIENT HAD RECTUM CANCER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE*ILS CURVED CIR STAPL | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Other| R |