FDA Adverse Event Injury Summary report: N

FISHER & PAYKEL HEALTHCARE

MDR report key: 18323768 · Received December 13, 2023

Report

Report Number
9611451-2023-01126
Event Type
Injury
Date Received
December 13, 2023
Report Date
November 15, 2023
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
UDI-DI
09420012407245
PMA / PMN Number
K110019
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THE COMPLAINT MR850 RESPIRATORY HUMIDIFIER IS CURRENTLY EN ROUTE TO FISHER & PAYKEL (F&P) HEALTHCARE FOR EVALUATION TO DETERMINE THE INVOLVEMENT OF OUR PRODUCT IN THE REPORTED EVENT. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

(B)(4). CORECTED DATA: D4 - SERIAL NUMBER AND UDI (UNIQUE DEVICE IDENTIFIER). D9 - DEVICE RETURNED TO MANUFACTURER - DATE OF RETURN. H6 - HEALTH EFFECT - IMPACT CODE. - MEDICAL DEVICE PROBLEM CODE. METHOD: THE SUBJECT MR850 RESPIRATORY HUMIDIFIER (SN (B)(6), THE 900MR805 HEATER WIRE ADAPTOR (LOT 2101452494) AND THE 900MR869 TEMPERATURE/FLOW PROBE (LOT 140407) IN USE DURING THE REPORTED EVENT WERE RETURNED TO THE FISHER & PAYKEL (F&P) HEALTHCARE IN NEW ZEALAND WHERE THEY WERE VISUALLY INSPECTED AND FUNCTIONALLY TESTED. OUR INVESTIGATION IS BASED ON THE INFORMATION PROVIDED BY THE HEALTHCARE FACILITY, F&P HEALTHCARE'S EVALUATION OF THE RETURNED DEVICES, AND OUR KNOWLEDGE OF THE PRODUCTS. RESULTS: VISUAL INSPECTION OF THE MR850 RESPIRATORY HUMIDIFIER IDENTIFIED SOME MINOR DAMAGE TO THE ENCLOSURE OF THE MR850 RESPIRATORY HUMIDIFIER. THERE WERE NO DEFECTS OR DAMAGE OBSERVED ON THE 900MR805 HEATER WIRE ADAPTOR OR THE 900MR869 TEMPERATURE/FLOW PROBE. THE MR850 RESPIRATORY HUMIDIFIER AND THE 900MR869 TEMPERATURE/FLOW PROBE PASSED FUNCTIONAL TESTING, INCLUDING THE ALARM SYSTEM CHECK. OTHER THAN THE ENCLOSURE DAMAGE, NO FAULTS OR MALFUNCTIONS WERE IDENTIFIED WITH EITHER DEVICE. DURING THE FUNCTIONAL TESTING OF THE 900MR805 HEATER WIRE ADAPTOR, AN INTERMITTENT HEATER WIRE ALARM OCCURRED DURING THE STARTUP OF THE SUBJECT MR850 RESPIRATORY HUMIDIFIER. FURTHER TESTING OF THE 900MR805 HEATER WIRE ADAPTOR CONFIRMED AN INTERMITTENT CONTINUITY BETWEEN ONE OF THE HEATER WIRE CONNECTOR PINS AND ITS WIRE. THE INTERMITTENT CONNECTION OF THE 900MR805 HEATER WIRE ADAPTOR WOULD NOT CAUSE OR CONTRIBUTE TO THE REPORTED ELEVATED TEMPERATURE ON THE MR850 RESPIRATORY HUMIDIFIER. A FAULTY HEATER WIRE ADAPTOR WOULD GENERATE AN ALARM CONDITION ON THE MR850 RESPIRATORY HUMIDIFIER, AND THE HUMIDIFIER WILL IMMEDIATELY TURN OFF POWER TO THE HEATER WIRES AND HEATER PLATE. THE HEALTHCARE FACILITY DID NOT CONFIRM THAT ANY ALARM CONDITIONS OCCURRED AT THE TIME OF THE REPORTED EVENT. CONCLUSION: BASED ON THE INFORMATION PROVIDED BY THE HEALTHCARE FACILITY AND THE INVESTIGATION WE HAVE UNDERTAKEN WE HAVE BEEN UNABLE TO DETERMINE THE CAUSE OF THE REPORTED TEMPERATURE FLUCTUATION ON THE MR850 RESPIRATORY HUMIDIFIER OR THE INJURY SUSTAINED BY THE PATIENT. THE SUBJECT MR850 RESPIRATORY HUMIDIFIER AND THE 900MR869 TEMPERATURE/FLOW PROBE WERE CONFIRMED TO BE OPERATING AS INTENDED AND WITHIN SPECIFICATION. THE MR850 RESPIRATORY HUMIDIFIER IS USED TO WARM AND HUMIDIFY GASES DELIVERED TO PATIENTS REQUIRING MECHANICAL VENTILATION, POSITIVE PRESSURE BREATHING ASSISTANCE OR MEDICAL GASES. THE MR850 COMPLIES WITH ISO8185:2007: "RESPIRATORY TRACT HUMIDIFIERS FOR MEDICAL USE - PARTICULAR REQUIREMENTS FOR RESPIRATORY HUMIDIFICATION SYSTEMS." THIS STANDARD SPECIFIES SAFETY REQUIREMENTS TO PREVENT THE RISK OF THERMAL INJURY, INCLUDING MAXIMUM ENTHALPY LIMITS FOR HUMIDIFIED GAS DELIVERED TO PATIENTS THROUGH RESPIRATORY HUMIDIFICATION SYSTEMS. THE MR850 RESPIRATORY HUMIDIFIER FEATURES AN AUDIBLE AND VISUAL ALARM WHICH ALERTS THE USER IF THE DISPLAYED TEMPERATURE EXCEEDS 41°C, OR IF THE AIRWAY TEMPERATURE EXCEEDS 43°C AND IMMEDIATELY DISABLES THE HEATER WIRE AND THE HEATER-PLATE. THE MR850 RESPIRATORY HUMIDIFIER USER INSTRUCTION STATES THE FOLLOWING WARNING: ENSURE THAT BOTH TEMPERATURE PROBE SENSORS ARE CORRECTLY AND SECURELY FITTED. FAILURE TO DO SO MAY RESULT IN TEMPERATURES IN EXCESS OF 41°C BEING DELIVERED TO THE PATIENT. THE RT265 INFANT EVAQUA2 BREATHING CIRCUIT USER INSTRUCTION ALSO STATES THE FOLLOWING WARNINGS: DO NOT USE HEATED WIRE BREATHING CIRCUITS WITHOUT GAS FLOW. IF GAS FLOW IS INTERRUPTED, TURN THE HUMIDIFIER OFF. AVOID PROLONGED CONTACT OF HEATED TUBES WITH PATIENT'S SKIN. CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE.

Additional Manufacturer Narrative · 0

(B)(4). **UDI RELATED DATA QUALITY UPDATES ONLY** CORRECTIONS TO REFLECT GUDID: SECTION D1: BRAND NAME UPDATED TO FISHER & PAYKEL HEALTHCARE. SECTION D2A: COMMON DEVICE NAME UPDATED TO RESPIRATORY HUMIDIFIER. SECTION D4: MODEL AND CATALOG # UPDATED TO MR850ALU. SECTION D4: UDI DETAILS ARE NOT AVAILABLE BASED ON THE TIME OF MANUFACTURING. SECTION G1: CONTACT PERSON UPDATED TO MR. (B)(4). METHOD: THE SUBJECT MR850 RESPIRATORY HUMIDIFIER (SN (B)(6)), THE 900MR805 HEATER WIRE ADAPTOR (LOT 2101452494) AND THE 900MR869 TEMPERATURE/FLOW PROBE (LOT 140407) IN USE DURING THE REPORTED EVENT WERE RETURNED TO THE FISHER & PAYKEL (F&P) HEALTHCARE IN NEW ZEALAND WHERE THEY WERE VISUALLY INSPECTED AND FUNCTIONALLY TESTED. OUR INVESTIGATION IS BASED ON THE INFORMATION PROVIDED BY THE HEALTHCARE FACILITY, F&P HEALTHCARE'S EVALUATION OF THE RETURNED DEVICES, AND OUR KNOWLEDGE OF THE PRODUCTS. RESULTS: VISUAL INSPECTION OF THE MR850 RESPIRATORY HUMIDIFIER IDENTIFIED SOME MINOR DAMAGE TO THE ENCLOSURE OF THE MR850 RESPIRATORY HUMIDIFIER. THERE WERE NO DEFECTS OR DAMAGE OBSERVED ON THE 900MR805 HEATER WIRE ADAPTOR OR THE 900MR869 TEMPERATURE/FLOW PROBE. THE MR850 RESPIRATORY HUMIDIFIER AND THE 900MR869 TEMPERATURE/FLOW PROBE PASSED FUNCTIONAL TESTING, INCLUDING THE ALARM SYSTEM CHECK. OTHER THAN THE ENCLOSURE DAMAGE, NO FAULTS OR MALFUNCTIONS WERE IDENTIFIED WITH EITHER DEVICE. DURING THE FUNCTIONAL TESTING OF THE 900MR805 HEATER WIRE ADAPTOR, AN INTERMITTENT HEATER WIRE ALARM OCCURRED DURING THE STARTUP OF THE SUBJECT MR850 RESPIRATORY HUMIDIFIER. FURTHER TESTING OF THE 900MR805 HEATER WIRE ADAPTOR CONFIRMED AN INTERMITTENT CONTINUITY BETWEEN ONE OF THE HEATER WIRE CONNECTOR PINS AND ITS WIRE. THE INTERMITTENT CONNECTION OF THE 900MR805 HEATER WIRE ADAPTOR WOULD NOT CAUSE OR CONTRIBUTE TO THE REPORTED ELEVATED TEMPERATURE ON THE MR850 RESPIRATORY HUMIDIFIER. A FAULTY HEATER WIRE ADAPTOR WOULD GENERATE AN ALARM CONDITION ON THE MR850 RESPIRATORY HUMIDIFIER, AND THE HUMIDIFIER WILL IMMEDIATELY TURN OFF POWER TO THE HEATER WIRES AND HEATER PLATE. THE HEALTHCARE FACILITY DID NOT CONFIRM THAT ANY ALARM CONDITIONS OCCURRED AT THE TIME OF THE REPORTED EVENT. CONCLUSION: BASED ON THE INFORMATION PROVIDED BY THE HEALTHCARE FACILITY AND THE INVESTIGATION WE HAVE UNDERTAKEN WE HAVE BEEN UNABLE TO DETERMINE THE CAUSE OF THE REPORTED TEMPERATURE FLUCTUATION ON THE MR850 RESPIRATORY HUMIDIFIER OR THE INJURY SUSTAINED BY THE PATIENT. THE SUBJECT MR850 RESPIRATORY HUMIDIFIER AND THE 900MR869 TEMPERATURE/FLOW PROBE WERE CONFIRMED TO BE OPERATING AS INTENDED AND WITHIN SPECIFICATION. THE MR850 RESPIRATORY HUMIDIFIER IS USED TO WARM AND HUMIDIFY GASES DELIVERED TO PATIENTS REQUIRING MECHANICAL VENTILATION, POSITIVE PRESSURE BREATHING ASSISTANCE OR MEDICAL GASES. THE MR850 COMPLIES WITH ISO8185:2007: "RESPIRATORY TRACT HUMIDIFIERS FOR MEDICAL USE - PARTICULAR REQUIREMENTS FOR RESPIRATORY HUMIDIFICATION SYSTEMS." THIS STANDARD SPECIFIES SAFETY REQUIREMENTS TO PREVENT THE RISK OF THERMAL INJURY, INCLUDING MAXIMUM ENTHALPY LIMITS FOR HUMIDIFIED GAS DELIVERED TO PATIENTS THROUGH RESPIRATORY HUMIDIFICATION SYSTEMS. THE MR850 RESPIRATORY HUMIDIFIER FEATURES AN AUDIBLE AND VISUAL ALARM WHICH ALERTS THE USER IF THE DISPLAYED TEMPERATURE EXCEEDS 41°C, OR IF THE AIRWAY TEMPERATURE EXCEEDS 43°C AND IMMEDIATELY DISABLES THE HEATER WIRE AND THE HEATER-PLATE. THE MR850 RESPIRATORY HUMIDIFIER USER INSTRUCTION STATES THE FOLLOWING WARNING: - ENSURE THAT BOTH TEMPERATURE PROBE SENSORS ARE CORRECTLY AND SECURELY FITTED. FAILURE TO DO SO MAY RESULT IN TEMPERATURES IN EXCESS OF 41°C BEING DELIVERED TO THE PATIENT. THE RT265 INFANT EVAQUA2 BREATHING CIRCUIT USER INSTRUCTION ALSO STATES THE FOLLOWING WARNINGS: DO NOT USE HEATED WIRE BREATHING CIRCUITS WITHOUT GAS FLOW. IF GAS FLOW IS INTERRUPTED, TURN THE HUMIDIFIER OFF. AVOID PROLONGED CONTACT OF HEATED TUBES WITH PATIENT'S SKIN. CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE.

Description of Event or Problem · 0

A HEALTHCARE FACILITY IN SPAIN REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT A NEONATE SUSTAINED A THIRD DEGREE BURN TO THE NOSE WHEN RECEIVING HUMIDIFICATION FROM A F&P HEALTHCARE MR850 RESPIRATORY HUMIDIFIER VIA A F&P HEALTHCARE RT265 INFANT EVAQUA2 BREATHING CIRCUIT AND A NASAL MASK MANUFACTURERED BY A THIRD PARTY. IT WAS ALSO REPORTED THAT THE PATIENT REQUIRED ADDITIONAL MEDICAL TREATMENT AND SURGERY AS A RESULT OF THE INJURY. THERE WERE NO FURTHER PATIENT CONSEQUENCES REPORTED.

Description of Event or Problem · 0

A HEALTHCARE FACILITY IN SPAIN REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A PATIENT SUSTAINED A FACIAL BURN TO THE NOSE WHEN RECEIVING HUMIDIFIED GASES WHILE USING A NON-F&P CPAP NASAL MASK AND A MR850 RESPIRATORY HUMIDIFIER. IT WAS FURTHER REPORTED THAT THE INJURY WAS A THIRD DEGREE BURN AND REQUIRED MEDICAL TREATMENT AND SURGERY. SUBSEQUENT TO THAT, THE PATIENT'S CONDITION HAS BEEN REPORTED AS STABLE WHILE RESIDING AT HOME.

Description of Event or Problem · 0

A HEALTHCARE FACILITY IN SPAIN REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A PATIENT SUSTAINED A FACIAL BURN TO THE NOSE WHEN RECEIVING HUMIDIFIED GASES WHILE USING A NON-F&P CPAP NASAL MASK AND A MR850 RESPIRATORY HUMIDIFIER. IT WAS FURTHER REPORTED THAT THE INJURY WAS A THIRD DEGREE BURN AND REQUIRED MEDICAL TREATMENT AND SURGERY. SUBSEQUENT TO THAT, THE PATIENT'S CONDITION HAS BEEN REPORTED AS STABLE WHILE RESIDING AT HOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1780652 FISHER & PAYKEL HEALTHCARE RESPIRATORY HUMIDIFIER BTT FISHER & PAYKEL HEALTHCARE LTD MR850ALU 120403 09420012407245

Patients

Seq Age Sex Outcome Treatment
1 1 DA Male Required Intervention F&P 900MR805 HEATER WIRE ADAPTOR| F&P 900MR805 HEATER WIRE ADAPTOR| F&P 900MR869 TEMPERATURE/FLOW PROBE| F&P 900MR869 TEMPERATURE/FLOW PROBE| F&P RT265 INFANT EVAQUA2 BREATHING CIRCUIT| F&P RT265 INFANT EVAQUA2 BREATHING CIRCUIT| F&P RT265 INFANT EVAQUA2 BREATHING CIRCUIT| GETINGE® SERVO -N VENTILATOR| GETINGE® SERVO -N VENTILATOR| GETINGE® SERVO -N VENTILATOR| HAMILTON MEDICAL® 281976 NCPAP GENERATOR| HAMILTON MEDICAL® 281976 NCPAP GENERATOR| HAMILTON MEDICAL® 281976 NCPAP GENERATOR| HAMILTON MEDICAL® MEDIN NASAL MASK| HAMILTON MEDICAL® MEDIN NASAL MASK| HAMILTON MEDICAL® MEDIN NASAL MASK