FDA Adverse Event Malfunction Summary report: N

HEARTSTRING III PROXIMAL SEAL 4.3 MM

MDR report key: 1832307 · Received September 3, 2010

Report

Report Number
2242352-2010-02667
Event Type
Malfunction
Date Received
September 3, 2010
Date of Event
July 27, 2010
Report Date
August 6, 2010
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: THE DEVICE WAS RETURNED TO MAQUET CARDIAC SURGERY ON (B)(4) 2010, FOR INVESTIGATION. VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED STUCK IN THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE BODY OF THE LOADER, UNDER THE WINDOW. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD BE ATTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III PROXIMAL SEAL WOULDN'T "SNAP;" IT WOULDN'T DEPLOY. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT EFFECTS. THE SURGEON'S NAME IS UNKNOWN. THE PRODUCT IS RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTRING III PROXIMAL SEAL 4.3 MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HSK-3043 25010520

Patients

Seq Age Sex Outcome Treatment
1 NA