FDA Adverse Event Malfunction Summary report: N

PKG, 1688 AIM 4K CAMERA HEAD, C-MOUNT

MDR report key: 18323062 · Received December 13, 2023

Report

Report Number
0002936485-2023-01247
Event Type
Malfunction
Date Received
December 13, 2023
Date of Event
November 17, 2023
Report Date
July 2, 2024
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
GCJ
UDI-DI
07613327418989
PMA / PMN Number
K160332
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 0

THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE. ALLEGED FAILURE: POOR QUALITY IMAGE OUTPUT. CONFIRMED FAILURE: CHD-A-CBK - CABLE KINKED; CHD-A-DRE DEFORMED REAR ENCLOSURE; CHD-A-DTH DAMAGED THREADS ON FRONT ENCLOSURE. PROBABLE ROOT CAUSE: CABLES, HDMI CABLES, CONNECTORS, DIGITAL BOARD, POWER SUPPLY, FILTER / FUSE, AC INLET BOARD, MAIN BOARD, TRANSITION BOARD, INTERMITTENCE BETWEEN TRANSITION BOARD AND CH CONNECTOR, SOFTWARE, CAMERA HEAD (SENSOR, SENSOR CABLE), COUPLER (DUST ON OPTICS, SCRATCHED OPTICS, BROKEN / SHIFTED LENSES, FOCUSING MECHANISM, ENDOCLAMP, ZOOM RING); PROBLEM TOGGLING LIGHT SOURCE, CONNECTED MONITORS, LEAKING, POOR GROUNDING (E.G CAMERA HEAD CONNECTION FROM CABLE TO TRANSITION BOARD.), NO/INCORRECT COMMUNICATION WITH LIGHT SOURCE, SOAKING CAP DISCONNECTION DURING STERILIZATION, ELECTROMAGNETIC INTERFERENCE (EMI) FROM RF COMMUNICATION, HF SURGICAL INSTRUMENTS, ESD, OR POWER SURGE; CYBERSECURITY ATTACK; PENDULUM CH INERTIAL MEASUREMENT UNIT (IMU); INCIDENT LIGHT FROM SURGICAL SCENE IS REFLECTED BY COUPLER/CH WINDOW; USE ERROR.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS PARTIAL IMAGE LOSS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS PARTIAL IMAGE LOSS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1780432 PKG, 1688 AIM 4K CAMERA HEAD, C-MOUNT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ STRYKER ENDOSCOPY-SAN JOSE 07613327418989

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown