PKG, 1688 AIM 4K CAMERA HEAD, C-MOUNT
Report
- Report Number
- 0002936485-2023-01247
- Event Type
- Malfunction
- Date Received
- December 13, 2023
- Date of Event
- November 17, 2023
- Report Date
- July 2, 2024
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- GCJ
- UDI-DI
- 07613327418989
- PMA / PMN Number
- K160332
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE. ALLEGED FAILURE: POOR QUALITY IMAGE OUTPUT. CONFIRMED FAILURE: CHD-A-CBK - CABLE KINKED; CHD-A-DRE DEFORMED REAR ENCLOSURE; CHD-A-DTH DAMAGED THREADS ON FRONT ENCLOSURE. PROBABLE ROOT CAUSE: CABLES, HDMI CABLES, CONNECTORS, DIGITAL BOARD, POWER SUPPLY, FILTER / FUSE, AC INLET BOARD, MAIN BOARD, TRANSITION BOARD, INTERMITTENCE BETWEEN TRANSITION BOARD AND CH CONNECTOR, SOFTWARE, CAMERA HEAD (SENSOR, SENSOR CABLE), COUPLER (DUST ON OPTICS, SCRATCHED OPTICS, BROKEN / SHIFTED LENSES, FOCUSING MECHANISM, ENDOCLAMP, ZOOM RING); PROBLEM TOGGLING LIGHT SOURCE, CONNECTED MONITORS, LEAKING, POOR GROUNDING (E.G CAMERA HEAD CONNECTION FROM CABLE TO TRANSITION BOARD.), NO/INCORRECT COMMUNICATION WITH LIGHT SOURCE, SOAKING CAP DISCONNECTION DURING STERILIZATION, ELECTROMAGNETIC INTERFERENCE (EMI) FROM RF COMMUNICATION, HF SURGICAL INSTRUMENTS, ESD, OR POWER SURGE; CYBERSECURITY ATTACK; PENDULUM CH INERTIAL MEASUREMENT UNIT (IMU); INCIDENT LIGHT FROM SURGICAL SCENE IS REFLECTED BY COUPLER/CH WINDOW; USE ERROR.
IT WAS REPORTED THAT THERE WAS PARTIAL IMAGE LOSS.
IT WAS REPORTED THAT THERE WAS PARTIAL IMAGE LOSS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1780432 | PKG, 1688 AIM 4K CAMERA HEAD, C-MOUNT | LAPAROSCOPE, GENERAL & PLASTIC SURGERY | GCJ | STRYKER ENDOSCOPY-SAN JOSE | 07613327418989 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |