FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 1832138 · Received September 13, 2010

Report

Report Number
3005075853-2010-05217
Event Type
Malfunction
Date Received
September 13, 2010
Date of Event
July 16, 2010
Report Date
July 16, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE FIRED THROUGH LOCKOUT THE ANALYSIS RESULTS OF THE (B)(4) FOUND THAT IT WAS RETURNED EMPTY AND WITH THE LOCK OUT MECHANISM BROKEN AS IT WAS FIRED THROUGH. THE INSTRUMENT IS DESIGNED TO LOCK OUT AFTER ALL THE CLIPS HAVE BEEN FIRED; THEREFORE A POTENTIAL CAUSE OF THE CUSTOMER REPORTED EXPERIENCE IS THE FIRING OF ALL OF THE CLIPS AND THE INSTRUMENT "WOULD NOT FIRE" (ACTIVATION OF THE LOCK OUT MECHANISM). THE INSTRUMENT HAS AN ORANGE INDICATOR THAT APPEARS ON THE TOP OF THE HANDLE AS A REFERENCE FOR THE USER AS TO THE QUANTITY OF CLIPS REMAINING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN OPEN PROCEDURE THE DEVICE JAMMED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE. THERE IS NO ADDITIONAL INFORMATION AVAILABLE.