MDR report key: 1832065 · Received September 3, 2010

Report

Report Number
2028159-2010-01640
Event Type
Malfunction
Date Received
September 3, 2010
Date of Event
July 6, 2010
Report Date
July 6, 2010
Manufacturer
ALCON -IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE FOOTSWITCH WAS RECEIVED AND INSPECTED. THE LEFT AND RIGHT WING COVERS AND REAR BUMPER WERE DAMAGED. THE HEEL REST AND BASE PLATE WERE WORN. THE HOUSING AND HEEL BASE WERE BROKEN. THE FOOTSWITCH WAS REPAIRED. THE FOOTSWITCH WAS CLEANED AND LUBRICATED. THE FOOTSWITCH WAS THEN TESTED AND MET ALL PRODUCT SPECS. THE ROOT CAUSE OF THE REPORTED PROBLEM IS UNK. THE ISSUE WAS UNABLE TO BE REPLICATED IN-HOUSE. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADD'L SIMILAR REPORTS FOR THE SYSTEM OR THE FOOTSWITCH. (B)(4).

Description of Event or Problem · 1

ADVERSE EVENT(S): "NO PT INJURY REPORTED" (NO CONSEQUENCES OR IMPACT TO PT). PRODUCT PROBLEM(S): "FOOTSWITCH NOT RESPONDING" (NO CODE AVAILABLE). THE NURSE REPORTED THE FOOTSWITCH STOPPED RESPONDING. THE INCISION HAD BEEN MADE, BUT SURGERY HAD NOT STARTED. THE FOOTSWITCH WAS SWITCHED OUT WITHOUT RESOLUTION. THE SYSTEM WAS THEN SWITCHED OUT AND THE CASE WAS COMPLETED. THERE WAS A 20 MINUTE DELAY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON -IRVINE TECHNOLOGY CENTER INFINITI OZIL NA

Patients

Seq Age Sex Outcome Treatment
1 UNK