FDA Adverse Event Malfunction Summary report: N

CONSTELLATION WITH LASER

MDR report key: 1832020 · Received September 3, 2010

Report

Report Number
2028159-2010-01639
Event Type
Malfunction
Date Received
September 3, 2010
Date of Event
June 23, 2010
Report Date
August 5, 2010
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K063583
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY SERVICE REPRESENTATIVE EXAMINED THE SYSTEM AND CONFIRMED THE PROBLEM REPORTED. THE FLUIDICS MODULE WAS REPLACED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY WHEN ADD'L REPORTABLE INFO BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

ADVERSE EVENT(S): "NO PT INJURY REPORTED" (NO CONSEQUENCES OR IMPACT TO PT). PRODUCT PROBLEM(S): "SYSTEM WOULD NOT PRIME" (FAILURE TO PRIME); "SYSTEM MESSAGE DISPLAYED" (DEVICE ERROR MESSAGE). THE CUSTOMER REPORTED A SYSTEM MESSAGE DISPLAYED AND THE SYSTEM WOULD NOT PRIME. THE SYSTEM WAS SWITCHED OUT TO PROCEED WITH THE SURGERY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSTELLATION WITH LASER UNIT, PHACOFRAGMENTATION HQC ALCON - IRVINE TECHNOLOGY CENTER CONSTELLATION NA

Patients

Seq Age Sex Outcome Treatment
1 UNK