FDA Adverse Event Malfunction Summary report: N

C-ARM

MDR report key: 1831904 · Received September 8, 2010

Report

Report Number
9617766-2010-00514
Event Type
Malfunction
Date Received
September 8, 2010
Date of Event
August 12, 2010
Report Date
September 8, 2010
Manufacturer
GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD (GE PRIVATE)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. POWER SUPPLY CABLES WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO DISPLAY AN IMAGE. NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 C-ARM FLUOROSCOPY XRAY JAA GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD (GE PRIVATE) STENOSCOPE

Patients

Seq Age Sex Outcome Treatment
1