PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 18318247 · Received December 13, 2023

Report

Report Number
2024168-2023-13918
Event Type
Injury
Date Received
December 13, 2023
Date of Event
October 17, 2023
Report Date
April 10, 2024
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
UDI-DI
08717648113154
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PE
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING. D9 - DEVICE AVAILABLE FOR EVALUATION UPDATED FROM YES TO: NO.

Additional Manufacturer Narrative · 0

D4: LOT NUMBER UPDATED FROM 3052441 TO 3032841. D9/H3: SUBSEQUENT TO FILING THE INITIAL REPORTS, THE DEVICE WAS RECEIVED FOR ANALYSIS. H6: COMPONENT CODE 4755 REMOVED; 3088, 562, 1028, 3126 ADDED. H6: TYPE OF INVESTIGATION CODE 4115 REMOVED; 10 ADDED. VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.

Description of Event or Problem · 0

THIS WAS REPORTED AS AN ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY USING THE PRE-CLOSE TECHNIQUE VIA A 9F SHEATH HOLE PRIOR TO A ABDOMINAL AORTIC ANEURYSM INTERVENTION PROCEDURE. REPORTEDLY, ALTHOUGH THE FIRST TWO STEPS WERE COMPLETED PROPERLY, THE SUTURE WAS NOT PRESENT WHEN THE PLUNGER WAS REMOVED IN STEP 3. MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1580385 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 3032841 08717648113154

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention