IDENTITY IMPRINT PS
Report
- Report Number
- 3004153240-2023-00041
- Event Type
- Malfunction
- Date Received
- December 13, 2023
- Date of Event
- November 21, 2023
- Report Date
- December 12, 2023
- Manufacturer
- CONFORMIS, INC
- Product Code
- JWH
- PMA / PMN Number
- K221404
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A COMPLAINT WAS RECEIVED ON 21 NOV 2023 (B)(4) FROM A CONFORMIS SALES REPRESENTATIVE STATING THAT THE TIBIAL TRAY IMPLANT WAS INCORRECT. THE PACKAGE OF THE IMPRINT TRAY LOT NUMBER 972242 SHOULD'VE CONTAINED A SIZE 7 PS IMPRINT TRAY IMPLANT BUT CONTAINED A SIZE 7 CR TRAY. THE ERROR WAS NOT IDENTIFIED UNTIL AFTER THE CR TIBIAL TRAY AND PS FEMORAL IMPLANTS WERE BOTH CEMENTED IN PLACE. A SOFT CLOSE WAS DONE WITH THE PLAN TO COMPLETE THE SURGERY THE NEXT DAY. THE MODIFIED PS TRIAL WAS LEFT IN THE PATIENT WITH THE SOFT CLOSE. THE SECOND SURGERY WAS CONDUCTED ON (B)(6) 2023, WITH A CS INSERT IMPLANTED. SECOND SURGERY WITH CS INSERT WAS SUCCESSFUL. THERE ARE SEVERAL ASSIGNABLE ROOT CAUSES FOR THIS FIELD COMPLAINT: HUMAN ERROR WHEN PROGRAMMING THE IMPRINT TRAY ENGRAVING PROGRAMS. FAILURE OF FOLLOWING THE VALIDATED AND CONTROLLED QC INSPECTION PROCESS IN (B)(4). HUMAN ERROR WHEN PRINTING THE INTEROFFICE LABEL PER(B)(4). SYSTEM ERROR OF NON-ACTIVE LABEL STAYING IN WORK ORDER IN SAP SYSTEM. VISUAL INSPECTION FAILURES OCCURRING IN QC AND IN THE CLEANROOM WHEN INSPECTING AND PACKAGING THIS IMPRINT CR TRAY WITH LABELING OF AN IMPRINT PS TRAY.
IDENTITY IMPRINT PS PRODUCT CONTAINED IDENTITY IMPRINT CR TIBIAL TRAY. PS FEMUR AND CR TIBIAL TRAY IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1581226 | IDENTITY IMPRINT PS | PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL | JWH | CONFORMIS, INC | 972242D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Unknown | Other |