FDA Adverse Event Malfunction Summary report: N

ENDOPATH XCEL BLUNT TIP TROCAR

MDR report key: 1831700 · Received November 16, 2007

Report

Report Number
1527736-2007-07698
Event Type
Malfunction
Date Received
November 16, 2007
Date of Event
October 24, 2007
Report Date
October 29, 2007
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED WITH THE INNER SEAL ASSEMBLY AND DUCKBILL DEFORMED AS THE OBTURATOR WAS RETURNED INSERTED THROUGH THE SLEEVE. THE SEAL SET CAUSED BY STERILIZATION WITH THE OBTURATOR INSERTED THROUGH THE SEAL GENERALLY INCREASES THE LEAK RATE. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP APPENDECTOMY PROCEDURE THE TROCAR WAS LEAKING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.