ENDOPATH XCEL BLUNT TIP TROCAR
Report
- Report Number
- 1527736-2007-07698
- Event Type
- Malfunction
- Date Received
- November 16, 2007
- Date of Event
- October 24, 2007
- Report Date
- October 29, 2007
- Manufacturer
- ETHICON ENDO SURGERY, INC. (CINCINNATI)
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED WITH THE INNER SEAL ASSEMBLY AND DUCKBILL DEFORMED AS THE OBTURATOR WAS RETURNED INSERTED THROUGH THE SLEEVE. THE SEAL SET CAUSED BY STERILIZATION WITH THE OBTURATOR INSERTED THROUGH THE SEAL GENERALLY INCREASES THE LEAK RATE. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAP APPENDECTOMY PROCEDURE THE TROCAR WAS LEAKING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.