FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CXE VOLUMETRICINFUSION PUMP

MDR report key: 1831549 · Received September 10, 2010

Report

Report Number
6000001-2010-03158
Event Type
Malfunction
Date Received
September 10, 2010
Date of Event
August 1, 2010
Report Date
August 4, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF FAILURE CODE 810:11, AND THE ROOT CAUSE WAS DETERMINED TO BE AN OUT OF CALIBRATION AIR IN LINE PRINTED CIRCUIT BOARD. THE AIR IN LINE PRINTED CIRCUIT BOARD WAS RECALIBRATED TO REPAIR THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE STAPLES WERE NOT FORMING CORRECTLY. UNKNOWN HOW THE CASE WAS COMPLETED. THERE WAS NO PATIENT CONSEQUENCE. THE DEVICE WAS DISCARDED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED FAILURE CODE 810:11 DURING BIOMEDICAL TESTING. REVIEW OF THE DEVICE EVENT HISTORY DETERMINED THAT THIS CONDITION INTERRUPTED DELIVERY. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. THE PUMP WAS CATEGORIZED AS A REMEDIATED PUMP WITH SOFTWARE VERSION 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CXE VOLUMETRICINFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1