FDA Adverse Event Malfunction Summary report: N

AU600 CHEMISTRY ANALYZER

MDR report key: 1831489 · Received September 10, 2010

Report

Report Number
2050012-2010-00724
Event Type
Malfunction
Date Received
September 10, 2010
Date of Event
August 26, 2010
Report Date
September 10, 2010
Manufacturer
BECKMAN COULTER MISHIMA K.K.
Product Code
JJE
PMA / PMN Number
K961274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER POWERED THE UNIT OFF UPON NOTICING THE SMOKE. A BCI FIELD SERVICE ENGINEER (FSE) INVESTIGATED THE SMOKE AND CONCLUDED THAT THE CPU OF AU600 PC TO BE DEFECTIVE. FSE INSTALLED NEW COMPUTER, CLEANED CUVETTE, CALIBRATED AND QC TESTED THE UNIT. FSE VERIFIED ALL REPAIRS PER ESTABLISHED PROCEDURES AND ALL RESULTS MEET PUBLISHED PERFORMANCE SPECIFICATIONS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO SMOKE BEING RELEASED FROM THE AU600 CHEMISTRY ANALYZER. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AU600 CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER MISHIMA K.K. N/A N/A

Patients

Seq Age Sex Outcome Treatment
1