ARCHITECT I2000SR
Report
- Report Number
- 3016438761-2023-00653
- Event Type
- Malfunction
- Date Received
- December 12, 2023
- Date of Event
- November 21, 2023
- Report Date
- December 21, 2023
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- JJE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
UPDATED POSTAL CODE E1 (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
A REVIEW OF TRACKING AND TRENDING DATA DID NOT IDENTIFY ANY RELATED TRENDS FOR THE PRODUCT FOR THE ISSUE. A REVIEW OF THE ARCHITECT I2000SR, SERIAL # (B)(6) SERVICE HISTORY, REVEALED NO ADDITIONAL ISSUES OF SPARKING (CHARRING/BURN) OBSERVED DUE TO A PART, REPORTED ON THE (B)(6). THE ARCHITECT SYSTEM OPERATIONS MANUAL CONTAINED ADEQUATE INFORMATION FOR THE TROUBLESHOOTING, REMOVAL, AND REPLACEMENT OF THE ARC PWR CRD NA250 (LIST NUMBER 08L13-01). THE REPLACEMENT OF THE ARC PWR CRD NA250 WAS THE ISSUE RESOLUTION. THE 2023 UL CERTIFICATION MEMO CONTAINS INFORMATION NOTING ABBOTT DIAGNOSTIC EQUIPMENT AND ACCESSORIES ARE CERTIFIED TO THE APPROPRIATE SAFETY STANDARDS, AND ADEQUATE PROTECTION IS PROVIDED FOR THE OPERATOR AGAINST ELECTRICAL SHOCK FROM THE EQUIPMENT. THE SPARKING (CHARRING/BURN) OBSERVED WAS LIMITED TO ARC PWR CRD NA250 (LIST NUMBER 08L13-01); THE SPARKING (CHARRING/BURN) DID NOT SPREAD TO OTHER PARTS OF THE MODULE. BASED ON THE AVAILABLE INFORMATION NO PRODUCT DEFICIENCY OF THE ARC PWR CRD NA250 (LIST NUMBER 08L13-01) OR ARCHITECT I2000SR PROCESSING MODULE, SERIAL NUMBER (B)(6), WAS IDENTIFIED.
THE CUSTOMER OBSERVED THE NO POWER TO SCC OF ARCHITECT I2000SR PROCESSING MODULE. THE CUSTOMER TRIED TO DISCONNECT POWER CABLE AND RECONNECT BACK ON. THE CUSTOMER HEAR NOISE FROM SCC OF ARCHITECT I2000SR PROCESSING MODULE. THE CUSTOMER NOTICED BURN MARK INSIDE THE ON CABLE CONNECTOR. THERE WAS NO REPORTED INJURY TO ANY USERS, NO PATIENT IMPACT, AND NO FACILITY DAMAGE.
THE CUSTOMER OBSERVED THE NO POWER TO SCC OF ARCHITECT I2000SR PROCESSING MODULE. THE CUSTOMER TRIED TO DISCONNECT POWER CABLE AND RECONNECT BACK ON. THE CUSTOMER HEAR NOISE FROM SCC OF ARCHITECT I2000SR PROCESSING MODULE. THE CUSTOMER NOTICED BURN MARK INSIDE THE ON CABLE CONNECTOR. THERE WAS NO REPORTED INJURY TO ANY USERS, NO PATIENT IMPACT, AND NO FACILITY DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1383955 | ARCHITECT I2000SR | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | ABBOTT LABORATORIES |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |