FDA Adverse Event Malfunction Summary report: N

ARCHITECT I2000SR

MDR report key: 18313381 · Received December 12, 2023

Report

Report Number
3016438761-2023-00653
Event Type
Malfunction
Date Received
December 12, 2023
Date of Event
November 21, 2023
Report Date
December 21, 2023
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPDATED POSTAL CODE E1 (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 0

A REVIEW OF TRACKING AND TRENDING DATA DID NOT IDENTIFY ANY RELATED TRENDS FOR THE PRODUCT FOR THE ISSUE. A REVIEW OF THE ARCHITECT I2000SR, SERIAL # (B)(6) SERVICE HISTORY, REVEALED NO ADDITIONAL ISSUES OF SPARKING (CHARRING/BURN) OBSERVED DUE TO A PART, REPORTED ON THE (B)(6). THE ARCHITECT SYSTEM OPERATIONS MANUAL CONTAINED ADEQUATE INFORMATION FOR THE TROUBLESHOOTING, REMOVAL, AND REPLACEMENT OF THE ARC PWR CRD NA250 (LIST NUMBER 08L13-01). THE REPLACEMENT OF THE ARC PWR CRD NA250 WAS THE ISSUE RESOLUTION. THE 2023 UL CERTIFICATION MEMO CONTAINS INFORMATION NOTING ABBOTT DIAGNOSTIC EQUIPMENT AND ACCESSORIES ARE CERTIFIED TO THE APPROPRIATE SAFETY STANDARDS, AND ADEQUATE PROTECTION IS PROVIDED FOR THE OPERATOR AGAINST ELECTRICAL SHOCK FROM THE EQUIPMENT. THE SPARKING (CHARRING/BURN) OBSERVED WAS LIMITED TO ARC PWR CRD NA250 (LIST NUMBER 08L13-01); THE SPARKING (CHARRING/BURN) DID NOT SPREAD TO OTHER PARTS OF THE MODULE. BASED ON THE AVAILABLE INFORMATION NO PRODUCT DEFICIENCY OF THE ARC PWR CRD NA250 (LIST NUMBER 08L13-01) OR ARCHITECT I2000SR PROCESSING MODULE, SERIAL NUMBER (B)(6), WAS IDENTIFIED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED THE NO POWER TO SCC OF ARCHITECT I2000SR PROCESSING MODULE. THE CUSTOMER TRIED TO DISCONNECT POWER CABLE AND RECONNECT BACK ON. THE CUSTOMER HEAR NOISE FROM SCC OF ARCHITECT I2000SR PROCESSING MODULE. THE CUSTOMER NOTICED BURN MARK INSIDE THE ON CABLE CONNECTOR. THERE WAS NO REPORTED INJURY TO ANY USERS, NO PATIENT IMPACT, AND NO FACILITY DAMAGE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED THE NO POWER TO SCC OF ARCHITECT I2000SR PROCESSING MODULE. THE CUSTOMER TRIED TO DISCONNECT POWER CABLE AND RECONNECT BACK ON. THE CUSTOMER HEAR NOISE FROM SCC OF ARCHITECT I2000SR PROCESSING MODULE. THE CUSTOMER NOTICED BURN MARK INSIDE THE ON CABLE CONNECTOR. THERE WAS NO REPORTED INJURY TO ANY USERS, NO PATIENT IMPACT, AND NO FACILITY DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1383955 ARCHITECT I2000SR ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES

Patients

Seq Age Sex Outcome Treatment
1 Unknown