FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 1830901 · Received September 10, 2010

Report

Report Number
2017865-2010-03254
Event Type
Injury
Date Received
September 10, 2010
Date of Event
July 30, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
PMAP950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. ONLY A PARTIAL LEAD WAS RETURNED, MEASURING 13CM FROM THE CONNECTOR PIN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM AFTER RECEIVING MULTIPLE INAPPROPRIATE SHOCKS. THE STORED EGMS SHOWED NOISE ON THE LEAD. LEAD IMPEDANCE, CAPTURE AND SENSING WERE STABLE. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1581/60 NA

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention