FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 1830895 · Received September 7, 2010

Report

Report Number
2953144-2010-01875
Event Type
Injury
Date Received
September 7, 2010
Date of Event
August 13, 2010
Report Date
August 16, 2010
Manufacturer
ABBOTT VASCULAR -REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P960043
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE INDICATED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Description of Event or Problem · 1

DEVICE ISSUE #2: NEEDLE-TO-CUFF MISS. TIME OF DEVICE ISSUE: DURING VESSEL CLOSURE. ADVERSE EVENT: FAILURE TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, WHEN THE NEEDLE PLUNGER WAS REMOVED, A CUFF MISS OCCURRED. THE PROGLIDE WAS REMOVED AND A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. HEMOSTASIS WAS ACHIEVED USING MANUAL COMPRESSION. THERE WERE NO REPORTED PT ADVERSE EFFECTS. THOUGH REQUESTED, NO ADD'L INFO WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM MGB ABBOTT VASCULAR -REDWOOD CITY NA 91019-6H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention PART# 12673-03, LOT# 91019-6H| DEVICE #1 PERCLOSE PROGLIDE