FDA Adverse Event
Injury
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 1830777
·
Received September 10, 2010
Report
- Report Number
- 2017865-2010-03282
- Event Type
- Injury
- Date Received
- September 10, 2010
- Date of Event
- June 10, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- PMAP950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT PRESENTED TO FOLLOW-UP WITH INCREASED HIGH VOLTAGE LEAD IMPEDANCE. ON (B)(6)2010, DURING SURGICAL PROCEDURE IT WAS NOTED THAT THE HELIX OF THE RV COIL WAS NOT EXTENDED PROPERLY. AFT ER REDEPLOYING THE HELIX, THE HVLI VALUES WERE NORMAL. THE LEAD REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1581 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Required Intervention |