FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 1830440
·
Received September 10, 2010
Report
- Report Number
- 2017865-2010-03442
- Event Type
- Injury
- Date Received
- September 10, 2010
- Date of Event
- June 7, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- PMAP950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED SEVERAL INAPPROPRIATE HV THERAPIES. INTERFERENCES COULD BE SEEN ON THE EGM. DURING THE SURGICAL PROCEDURE, INSULATION DAMAGE AT THE TRIFURCATED CONNECTOR WAS OBSERVED, POSSIBLY DUE TO THE TIGHT LIGATURE NEAR AND UNDER THE SUTURE SLEEVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7121/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention |