FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 1830435
·
Received September 10, 2010
Report
- Report Number
- 2017865-2010-03451
- Event Type
- Injury
- Date Received
- September 10, 2010
- Date of Event
- June 30, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- PMAP950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD HAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7121/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention |