FDA Adverse Event
Malfunction
Summary report: N
E SERIES DEFIBRILLATOR
MDR report key: 1830414
·
Received September 1, 2010
Report
- Report Number
- 1220908-2010-02404
- Event Type
- Malfunction
- Date Received
- September 1, 2010
- Report Date
- August 13, 2010
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ZOLL MEDICAL CORP HAS NOT RECEIVED THE PRODUCT FOR EVAL, AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED UP WITHOUT A DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | E SERIES DEFIBRILLATOR | E SERIES | MKJ | ZOLL MEDICAL CORPORATION | E SERIES | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |