ULTRACISION HARMONIC SCALPEL HAND PIECE
MDR report key: 1830386
·
Received January 11, 2008
Report
- Report Number
- 1527736-2008-00234
- Event Type
- Malfunction
- Date Received
- January 11, 2008
- Date of Event
- December 14, 2007
- Report Date
- December 18, 2007
- Manufacturer
- ETHICON ENDO SURGERY, INC. (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVAL SUMMARY: THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT AND THE NOSE CONE CRACKED. THE HAND PIECE WAS DISASSEMBLED, A COLLAPSED/TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. THIS COULD HAVE BEEN THE CAUSE OF HOT HAND PIECE. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MFG PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A LAP GASTRIC BYPASS THE HANDPIECE IS GETTING HOT TO THE TOUCH. THE CUSTOMER REPLACED THE DEVICE AND COMPLETED THE CASE WITH NO PATIENT CONSEQUENCE.