FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1830386 · Received January 11, 2008

Report

Report Number
1527736-2008-00234
Event Type
Malfunction
Date Received
January 11, 2008
Date of Event
December 14, 2007
Report Date
December 18, 2007
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL SUMMARY: THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT AND THE NOSE CONE CRACKED. THE HAND PIECE WAS DISASSEMBLED, A COLLAPSED/TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. THIS COULD HAVE BEEN THE CAUSE OF HOT HAND PIECE. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP GASTRIC BYPASS THE HANDPIECE IS GETTING HOT TO THE TOUCH. THE CUSTOMER REPLACED THE DEVICE AND COMPLETED THE CASE WITH NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE LFL ETHICON ENDO SURGERY, INC. (CINCINNATI) NA NA

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR