MAGNESIUM
Report
- Report Number
- 3005094123-2023-00364
- Event Type
- Malfunction
- Date Received
- December 11, 2023
- Date of Event
- December 1, 2023
- Report Date
- February 22, 2024
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- JGJ
- UDI-DI
- 00380740176532
- PMA / PMN Number
- K181748
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ALINITY C MAGNESIUM, LIST 08P19-34, ABBOTT IRELAND DIAGNOSTICS DIVISION, LONGFORD, IRELAND TO THE ALINITY C PROCESSING MODULE, LIST 03R67-01, ABBOTT LABORATORIES, IRVING, TEXAS, USA. MDR NUMBER 3016438761-2024-00113-00 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. COMPLETE INFORMATION FOR SECTION A1 PATIENT IDENTIFIER IS (B)(6).
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C MAGNESIUM RESULTS FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED WHICH WAS NOT REPORTED OUT OF THE LAB: SID (B)(6)PROCESSED ON (B)(6)2023 INITIAL MAGNESIUM RESULT = 3.44 MMOL/L REPEAT 0.53 MMOL/L REFERENCE RANGE = 0.66-1.07 MMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C MAGNESIUM RESULTS FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED WHICH WAS NOT REPORTED OUT OF THE LAB: SID (B)(6) PROCESSED ON (B)(6) 2023 INITIAL MAGNESIUM RESULT = 3.44 MMOL/L REPEAT 0.53 MMOL/L REFERENCE RANGE = 0.66-1.07 MMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2180550 | MAGNESIUM | PHOTOMETRIC METHOD, MAGNESIUM | JGJ | ABBOTT IRELAND DIAGNOSTICS DIVISION | 52268UD00 | 00380740176532 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ALNTY C PROCESSING MODU, 03R67-01, (B)(6)| ALNTY C PROCESSING MODU, 03R67-01, (B)(6) |