FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 1830095 · Received September 7, 2010

Report

Report Number
9615739-2010-00211
Event Type
Malfunction
Date Received
September 7, 2010
Date of Event
August 12, 2010
Report Date
August 12, 2010
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE BAR WAS NOT ENGAGING THE BRAKE SET SENSOR AND WHEN HE MOVED THE BED AROUND, THE RIGHT FOOT END CASTER MOVED MORE THAN IT SHOULD WHEN IN BRAKE. THE TECH FOUND THE CASTER SUPPORT WAS CRACKED IN HALF. THE TECH REPLACED THE CASTER SUPPORT TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE BRAKE IS NOT SET AND THE BED IS ALARMING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC POWERED HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1