FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1830086 · Received September 7, 2010

Report

Report Number
3004209178-2010-82711
Event Type
Malfunction
Date Received
September 7, 2010
Date of Event
July 31, 2010
Report Date
August 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A FROZEN DISPLAY AS A RESULT OF CORROSION ON THE LIQUID CRYSTAL DISPLAY BOARD AND THE KEYPAD TRACE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP WAS NOT FUNCTIONING AFTER THE DEVICE WAS DROPPED IN THE TOILET. THE CUSTOMER STATED THAT SHE DRIED THE DEVICE, BUT THE BUTTONS WERE NOT RESPONDING. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 36 YR