FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-722NAS PRDGM INS V2.2 SK EN
MDR report key: 1830086
·
Received September 7, 2010
Report
- Report Number
- 3004209178-2010-82711
- Event Type
- Malfunction
- Date Received
- September 7, 2010
- Date of Event
- July 31, 2010
- Report Date
- August 13, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A FROZEN DISPLAY AS A RESULT OF CORROSION ON THE LIQUID CRYSTAL DISPLAY BOARD AND THE KEYPAD TRACE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP WAS NOT FUNCTIONING AFTER THE DEVICE WAS DROPPED IN THE TOILET. THE CUSTOMER STATED THAT SHE DRIED THE DEVICE, BUT THE BUTTONS WERE NOT RESPONDING. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAS PRDGM INS V2.2 SK EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |