FDA Adverse Event Death Summary report: N

INFUSOMAT®

MDR report key: 18300534 · Received December 11, 2023

Report

Report Number
2523676-2023-00818
Event Type
Death
Date Received
December 11, 2023
Report Date
November 6, 2024
Manufacturer
B. BRAUN MEDICAL INC.
Product Code
FPA
UDI-DI
04046964186127
PMA / PMN Number
K142036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). NO SAMPLE WAS PROVIDED FOR EVALUATION. DUE TO COMPLAINTS OF THIS NATURE, CORRECTIVE ACTIONS WERE PERFORMED TO REVISE SPECIFICATIONS WHICH LIMIT UNIT-TO-UNIT VARIATION IN TUBING OUTER DIAMETERS AND INCREASED THE LENGTH OF THE TUBING THAT CONTACTS THE AIR SENSOR. THE REDUCED VARIATION IN TUBING OUTER DIAMETER AND LENGTH REVISION IMPROVES THE COUPLING BETWEEN THE ADMINISTRATION SET AND THE AIR-IN-LINE SENSOR REDUCING THE POTENTIAL FOR FALSE AIR-IN-LINE ALARMS. B. BRAUN MEDICAL INC. HAS INITIATED A VOLUNTARY MEDICAL DEVICE CORRECTION 2523676-9/26/23-004-C FOR MULTIPLE BATCHES OF INFUSOMAT® PUMP SETS MANUFACTURED BETWEEN 01.JAN.2022 AND 17.AUG.2023. WE WILL MAINTAIN THIS REPORT FOR FURTHER REFERENCES AND CONTINUE TO MONITOR OTHER REPORTS FOR SIMILAR OCCURRENCES. IF ANY ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE, A FOLLOW UP WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW UP WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.

Description of Event or Problem · 0

AS PER MEDWATCH MW5148159: "PATIENT HAD LAPAROSCOPIC SURGERY WITH SMALL INTESTINE RESECTION ON (B)(6) 2023. ON (B)(6) 2023 AT 2355 PATIENT BECAME HYPOTENSIVE (BP 75/50), TACHYCARDIC (HR 20'S), AND TACHYPNEIC (RR HIGH-40'S). RAPID RESPONSE CALLED AND THE PATIENT WAS TRANSFERRED TO THE ICU. UTILIZING THE BBRAUN INFUSOMAT SPACE PUMPS THE PATIENT WAS STARTED ON LEVOPHED, AND THEN VASOPRESSIN AND NEOSYNEPHRINE ARE ADDED. THE BBRAUN ADMINISTRATION SETS UTILIZED (INFUSOMAT SPACE 15 DRIP/ML 120 INCH TUBING) ARE THE SUBJECT OF A BBRAUN URGENT MEDICAL DEVICE CORRECTION NOTICE DATED OCTOBER 23, 2023 AND THIS NOTIFICATION WAS RECEIVED VIA US MAIL AT THE HOSPITAL ON NOVEMBER 6, 2023. THERE ARE NO UNAFFECTED TUBING ADMINISTRATION LOTS IN COMMERCIAL CIRCULATION AT THIS TIME WITH NEW IMPROVED INVENTORY DUE ON NOVEMBER 30, 2023 PER THE BBRAUN DEVICE CORRECTION NOTICE. AS NOTED IN THE CORRECTION NOTICE BBRAUN HAS IDENTIFIED THAT THEIR OUTER DIAMETER TUBING MAY VARY IN SIZE LEADING TO INCREASED AIR ALARMS, INCLUDING FALSE AIR ALARMS AS WELL AS THE CREATION OF MICROBUBBLES IN THE TUBING ADMINISTRATION SETS, FURTHER LEADING TO ADDITIONAL AIR ALARMS. DUE TO MULTIPLE AIR ALARMS GENERATED BY THE BBRAUN INFUSOMAT SPACE PUMPS IN USE (8 MULTIPLE BBRAUN INFUSOMAT SPACE PUMPS RUNNING FOR THIS CRITICALLY ILL PATIENT) AS A RESULT OF THE ADMINISTRATION SET ISSUE THIS PATIENT EXPERIENCED FURTHER INSTABILITY LEADING TO THE INTERRUPTION OF THE INFUSION OF VASOPRESSORS, LEVOPHED AND NEOSYNEPHRINE IN A HEMODYNAMICALLY UNSTABLE PATIENT. THE PATIENT HAS SUBSEQUENTLY PASSED AWAY ON (B)(6) 2023. AS SUCH WE ARE REPORTING THIS EVENT TO THE FDA. REFERENCE REPORTS: MW5148158, MW5148160."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1845614 INFUSOMAT® SET, ADMINISTRATION, INTRA FPA B. BRAUN MEDICAL INC. 0061891636 04046964186127

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Life Threatening| D