FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 1830053 · Received September 7, 2010

Report

Report Number
9615739-2010-00212
Event Type
Malfunction
Date Received
September 7, 2010
Date of Event
August 13, 2010
Report Date
August 13, 2010
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE STEERING CASTER WAS DAMAGED AND FOUND TWO CASTER SUPPORTS WERE BROKEN. THE TECH REPLACED THE STEER AND BRAKE CASTER AND THE TWO CASTER SUPPORTS TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BRAKE IS NOT HOLDING OR STEERING PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC POWERED HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1