SYNCHRON® CX5CE CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2010-00734
- Event Type
- Malfunction
- Date Received
- September 9, 2010
- Date of Event
- August 19, 2010
- Report Date
- September 9, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K881495
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ONTO LOCATION AND IDENTIFIED THE ROOT CAUSE OF THE LEAK TO BE IMPROPERLY SEALED SAMPLE AND REAGENT SYRINGES INTO THE 3-WAY T-VALVE. FSE TIGHTENED THE SYRINGES, WHICH RESOLVED THE ISSUE. UPON REOCCURRENCE, THE HARDWARE FAILURES COULD CAUSE ERRONEOUS RESULTS ON ANY CHEMISTRIES AND PATIENT TREATMENT AND/OR OPERATOR SAFETY IS LIKELY TO BE IMPACTED.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING A LEAK FROM THE HYDROPNEUMATICS THAT WAS CAUSED BY LOOSE SAMPLE AND REAGENT SYRINGES IN THE SYNCHRON CX5CE CHEMISTRY ANALYZER. THIS EVENT COULD POTENTIALLY CAUSE A SHORT SAMPLE/REAGENT DELIVERY OR DILUTE THE SAMPLE/REAGENT DUE TO THE DRIPPING. NO PATIENT RESULTS WERE GENERATED DURING THIS EVENT, THEREFORE, NO TREATMENT WAS IMPACTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® CX5CE CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | CX5CE | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |