FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2010-01252
- Event Type
- Malfunction
- Date Received
- September 9, 2010
- Date of Event
- August 19, 2010
- Report Date
- October 6, 2010
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
CUSTOMER'S METER AND TEST STRIP LOT # 1014117 WERE RETURNED AND INVESTIGATED. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE NOT FOUND IN METER MEMORY. THIS IS A FINAL REPORT.
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE FILED ONCE INVESTIGATION RESULTS ARE AVAILABLE.
IT WAS REPORTED TO BAXTER (B)(4) THAT THE RESERVOIR OF ONE (1) CE INFUSOR LV 2ML/HR DEVICE HAD RUPTURED DURING FILLING. THERE IS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 72 MG/DL AND 372 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.