CUSA NXT CONSOLE (115-230V)
Report
- Report Number
- 1222895-2010-00012
- Event Type
- Malfunction
- Date Received
- August 27, 2010
- Date of Event
- April 26, 2010
- Report Date
- August 27, 2010
- Manufacturer
- INTEGRA RADIONICS, INC.
- Product Code
- LFL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
THE NXT CONSOLE AND SERVICE MODULE WAS EVALUATED. THE UNIT WAS TURNED ON AND OPERATED WITH TWO SEPARATE 35 KHZ HAND PIECES WITH NO SYSTEM ERRORS OBSERVED. ROOT CAUSE IS INDETERMINATE AT THIS TIME AS THE ERROR OBSERVED AT THE SITE COULD NOT BE REPLICATED. ALL TESTING PERFORMED ON THE NXT UNIT INDICATES THAT THE UNIT IS IN GOOD WORKING ORDER. IT WAS NOTED THAT WHEN THE INITIAL ERROR WAS OBSERVED ON THE NXT, THE CANISTER LINER WAS CHANGED CORRECTING THE INSUFFICIENT SUCTION. ERRORS OBSERVED AT THE SITE COULD HAVE BEEN TRIGGERED BY A BAD CONNECTION ON THE CANISTER WHICH CAUSED THE UNIT TO NOT FUNCTION PROPERLY. A DHF REVIEW WAS PERFORMED AND NO INCONSISTENCIES WERE IDENTIFIED. INTEGRA CONSIDERS THIS COMPLAINT INVESTIGATION CLOSED. FUTURE INCIDENTS OF THIS NATURE WILL BE DOCUMENTED FOR RECURRENCE AND TRENDING PURPOSES.
ON (B)(6) 2010, A LIVER RESECTION WAS BEING PERFORMED ON A PATIENT. AFTER FIFTEEN PLUS MINUTES OF ULTRASONIC USE OF THE CUSA NXT, THE SUCTION DECREASED TO 14MMHG AND HAD AN ERROR OF INSUFFICIENT SUCTION. THE SCRUB NURSE FLUSHED THE HAND PIECE MULTIPLE TIMES BUT THERE WAS NO CHANGE. THE CANISTER LINER WAS CHANGED AND THIS RESOLVED THE SUCTION PROBLEM. THE SURGEON THEN BEGAN USING THE CUSA NXT AGAIN AND HE HAD NO ULTRASONIC POWER. THERE WAS NO ALARM AT FIRST AND THEN THE POWER ALARM - CAME OFF WITH E0003 CODE. THERE WAS NO PATIENT INJURY. THE DELAY IN SURGERY WAS REPORTED AS LESS THAN ONE MINUTE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CUSA NXT CONSOLE (115-230V) | ULTRASONIC SURGICAL PRODUCTS | LFL | INTEGRA RADIONICS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |