MDR report key: 18290630 · Received December 8, 2023

Report

Report Number
3023140882-2023-00001
Event Type
Death
Date Received
December 8, 2023
Date of Event
November 10, 2023
Report Date
December 22, 2023
Manufacturer
QUANTUM SURGICAL SAS
Product Code
JAK
PMA / PMN Number
K223758
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE TECHNICAL INVESTIGATION BASED ON EPIONE DEVICE LOGS, AND COLLECTED CT SCANS, SHOW THAT THERE IS NO EVIDENCE OF DEVICE MALFUNCTION, AND THE DEVICE HAS BEEN CONFIRMED TO BE WITHIN SPECIFICATIONS FOR ACCURACY BY TESTING PERFORMED WITH THE SAME SET OF INSTRUMENTS AS USED DURING THE INTERVENTION. BASED ON THE INVESTIGATION PERFORMED, THE MOST PROBABLE ROOT CAUSE IS THAT THE PLANNED NEEDLE TRAJECTORIES WERE TOO CLOSE AND/OR CROSSING CRITICAL STRUCTURES SUCH AS VESSELS. OVER THE 6 PLANNED NEEDLES TRAJECTORIES, 5 OF THEM WERE CLOSE TO ANATOMICAL STRUCTURES AT RISK SUCH AS VESSELS AND 2 OF THE 5 WERE CROSSING VESSELS. IT WAS NOTED THAT NO INTRAOPERATIVE CONTROL SCAN WAS PERFORMED AFTER COMPLETING NEEDLE INSERTION AS RECOMMENDED PER USER MANUAL (D-SY-DOU-0860A_FR QUICK_REFERENCE_GUIDE). THE COMBINATION OF THE LITTLE TO NO MARGIN VS CRITICAL STRUCTURES OF THE PLANNED TRAJECTORIES, AND BIOMECHANICAL EFFECTS THAT HAVE NOT BEEN DETECTED BEFORE END OF THE INSERTION, DUE TO THE ABSENCE OF INTERMEDIATE CONTROL CT SCAN ARE THE MAIN ROOT CAUSES OF THE ADVERSE EVENT OBSERVED DURING THE PROCEDURE.

Description of Event or Problem · 0

IT AS REPORTED TO QUANTUM SURGICAL ON THE 10 NOVEMBER 2023 THAT DURING AN INTERVENTION, A PATIENT TREATED FOR LIVER CANCER BY IRE (IRREVERSIBLE ELECTROPORATION) WITH EPIONE DEVICE TO PLACE 6 NEEDLES, HAD A SEVERE ABDOMEN BLEEDING AT THE END OF THE INTERVENTION. PATIENT WAS SENT TO ANGIOGRAPHIC ROOM TO HAVE ARTERIES EMBOLIZATION AND HAVE BLOOD TRANSFUSION. IT WAS REPORTED THAT PATIENT CONDITION WAS STABLE ON THE 10 NOVEMBER 2023. ADDITIONAL INFORMATION WAS COMMUNICATED TO QUANTUM SURGICAL ON THE 13TH OF NOVEMBER 2023 TO INFORM US THAT PATIENT HAD PASSED ON THE (B)(6) 2023.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
221363 EPIONE EPIONE JAK QUANTUM SURGICAL SAS EPIONE

Patients

Seq Age Sex Outcome Treatment
1 70 YR Male Other| D