LVP PRIMARY ADMIN SET 2 INLET
Report
- Report Number
- 3014732157-2023-00200
- Event Type
- Malfunction
- Date Received
- December 7, 2023
- Date of Event
- November 7, 2023
- Report Date
- August 21, 2024
- Manufacturer
- FRESENIUS KABI USA, LLC
- Product Code
- FPA
- PMA / PMN Number
- K183311
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE FOLLOWING HAS BEEN REPORTED DURING TRAINING BY A FRESENIUS KABI TRAINER: "PUMPS ARE ALARMING SECONDARY OCCLUSION ALARM WHEN TRYING SYRINGES (MEDLINE 20 ML SYRINGES SYR12001) OR SECONDARY LINES (BBRAUN SECONDARY SET V1921 40 IN). I HAVE SWAPPED OUT 10 PUMPS AND 10 SETS AND I TRIED DIFFERENT SYRINGES AND DIFFERENT SECONDARY SET, ALL HAVE THE SAME OCCLUSION ALARM. I HAVE TRIED TAKING THE SET OFF AND PUTTING IT BACK ON, I HAVE TRIED PULLING THE SYRINGE BARREL OUT AND BACK TO BREAK THE STICTION. WHEN I USED MY SET I BROUGHT FROM HOME WITH A DIFFERENT LOT- MINE WORKED. THE SECONDARY SET AND THE SYRINGE BOTH WORKED ON MY SET THAT I BROUGHT WITH ME." REPORTING DUE TO THE REFERENCED ISSUE. OCCURRED DURING TRAINING; NO ADVERSE EFFECTS WERE REPORTED. MORE INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.
EVALUATION STEPS PERFORMED: 1. VISUAL INSPECTION. 2. INSTALLED THE KIT ON IVENIX INFUSION SYSTEM MACHINE AND RAN THE PRIMING PROCESS. 3. UNDERWATER LEAK TEST. 255 SAMPLES PACKED WERE RETURNED FOR EVALUATION IN 11 BOXES. DURING THE VISUAL INSPECTION IT WAS OBSERVED THAT THE ASSEMBLIES OF THE KIT WERE ACCORDING TO THE DRAWING. A SLIT WAS OBSERVED VISIBLE AT THE LUER ACTIVATED VALVE, STRAIGHT (SECONDARY PORT). THE SAMPLE COMPLETED THE PRIMARY BOLUS PRIME AND SECONDARY PRIME PROCESS SUCCESSFULLY, THE REPORTED ALARM WAS NOT REPLICATED. THE UNIT BACK PRIME ON SECONDARY PORT WAS PERFORMED CORRECTLY. NO BLOCKAGE WAS DETECTED AT THE KIT AND SECONDARY PORT DURING EVALUATION PERFORMED. NO LEAKS WERE FOUND DURING THE UNDERWATER LEAK TEST. NO MANUFACTURING DEFECTS WERE FOUND IN THE SAMPLES RECEIVED. THE CUSTOMER COMPLAINT IS NOT CONFIRMED. OTHER SAMPLE EVALUATIONS: VISUALLY CONFIRMED ON THE SURFACE OF THE I-SITE LAV THAT THERE WAS NO SLIT THROUGH THE TOP SURFACE OF THE SILICONE. THEREFORE, THE FLUID WAS BLOCKED FROM MOVING IN EITHER DIRECTION THROUGH THE VALVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 220676 | LVP PRIMARY ADMIN SET 2 INLET | ADMINISTRATION SET | FPA | FRESENIUS KABI USA, LLC | 3006073 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |