FDA Adverse Event Malfunction Summary report: N

X-TACK¿ ENDOSCOPIC HELIX TACKING SYSTEM

MDR report key: 18286120 · Received December 7, 2023

Report

Report Number
3006722112-2023-00231
Event Type
Malfunction
Date Received
December 7, 2023
Date of Event
January 17, 2022
Report Date
October 9, 2023
Manufacturer
APOLLO ENDOSURGERY, INC
Product Code
OCW
PMA / PMN Number
K201808
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

COMBINED MEDWATCH SUBMITTED TO THE FDA ON 07/DEC/2023. CLINICAL STUDY: (B)(6). A REVIEW OF THE DEVICE LABELING NOTES THE FOLLOWING: THE CURRENT X-TACK¿ ENDOSCOPIC HELIX TACKING SYSTEM INSTRUCTIONS FOR USE (IFU) ADDRESSED THE KNOWN AND POTENTIAL EVENTS OF "BLEEDING " AS FOLLOWS: WARNINGS: - THE DEVICE SHOULD NOT BE USED TO TREAT ACUTELY BLEEDING ULCERS, ULCERS WITH STIGMATA OF RECENT BLEEDING OR ANY ULCERS WITH A VISIBLE VESSEL. - DO NOT USE A DEVICE WHERE THE INTEGRITY OF THE STERILE PACKAGING HAS BEEN COMPROMISED OR IF THE DEVICE APPEARS DAMAGED. - ONLY PHYSICIANS POSSESSING SUFFICIENT SKILL AND EXPERIENCE IN SIMILAR OR THE SAME TECHNIQUES SHOULD PERFORM ENDOSCOPIC PROCEDURES. - CONTACT OF ELECTROSURGICAL COMPONENTS WITH OTHER COMPONENTS MAY RESULT IN INJURY TO THE PATIENT AND/OR OPERATOR AS WELL AS DAMAGE TO THE DEVICE AND/OR ENDOSCOPE. - VERIFY COMPATIBILITY OF ENDOSCOPE SIZE, ENDOSCOPIC INSTRUMENTS AND ACCESSORIES AND ENSURE PERFORMANCE IS NOT COMPROMISED. - ENSURE ENDOSCOPE IS CLEAN, DRY, AND FREE OF LUBRICANTS PRIOR TO DEVICE INSTALLATION. - ENSURE ALL ENDOSCOPIC SCOPES, INCLUDING SCOPE CHANNELS, ARE IN GOOD WORKING CONDITION PRIOR TO USE. - SUCTION OPERATION THROUGH ENDOSCOPE MAY BE SIGNIFICANTLY REDUCED WHEN THE SCOPE CHANNEL LINER IS IN PROPER POSITION. - DO NOT PUSH THROUGH OR PULL BACK ON A RETROFLEXED SCOPE WITH INSTALLED HELIX TACK. - APPLYING EXCESSIVE FORCE TO THE DISTAL END OF THE X-TACK¿ DEVICE COULD COMPRESS OR DAMAGE THE HELIX TACK WHEN INSTALLED. - DO NOT RETRACT DEVICE INTO SCOPE WHILST A HELIX TACK IS INSTALLED. - REUSE OR REPROCESSING OF THE X-TACK¿ SYSTEM COULD RESULT IN DEVICE MALFUNCTION, PRECAUTIONS: - THE SYSTEM MAY ONLY BE USED IF PURCHASED FROM APOLLO ENDOSURGERY, INC. OR ONE OF ITS AUTHORIZED AGENTS. ADVERSE EVENTS: POSSIBLE COMPLICATIONS THAT MAY RESULT FROM USING THE X-TACK¿ SYSTEM INCLUDE, BUT MAY NOT BE LIMITED TO: - PHARYNGITIS / SORE THROAT. - NAUSEA AND / OR VOMITING. - ABDOMINAL PAIN AND / OR BLOATING. - HEMORRHAGE. - HEMATOMA. - CONVERSION TO LAPAROSCOPIC OR OPEN PROCEDURE. - STRICTURE. - INFECTION / SEPSIS. - PHARYNGEAL, GASTRIC, COLONIC AND/OR ESOPHAGEAL PERFORATION. - ESOPHAGEAL, GASTRIC, COLONIC AND/OR PHARYNGEAL LACERATION. - INTRA-ABDOMINAL (HOLLOW OR SOLID) VISCERAL INJURY. - ASPIRATION. - WOUND DEHISCENCE. - ACUTE INFLAMMATORY TISSUE REACTION. - DEATH. ADDITIONAL INFORMATION: THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. DHR REVIEW WAS COMPLETED FOR LOT NUMBER, AF04669. THE SUBJECT PRODUCT MET ALL SPECIFICATIONS AND REQUIREMENTS IN EFFECT AT THE TIME OF MANUFACTURE. THERE WAS ONE OTHER COMPLAINT IN THE APOLLO DATABASE AGAINST THIS LOT NUMBER AF04669. ADDITIONAL INFORMATION: DEVICE EVALUATION SUMMARY: ASSESSMENT OF THE DEVICE INVOLVED IN THIS COMPLAINT WAS NOT POSSIBLE, AND IT HAS NOT BEEN POSSIBLE TO DETERMINE THE ROOT CAUSE FOR THIS EVENT.

Description of Event or Problem · 0

CLINICAL STUDY. IT WAS REPORTED BLEEDING DUE TO SUTURE ON THE FOLLOW UP VISIT 11 DAYS LATER. SUTURING DURING ENDOSCOPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259159 X-TACK¿ ENDOSCOPIC HELIX TACKING SYSTEM ENDOSCOPIC HELIX TACKING SYSTEM OCW APOLLO ENDOSURGERY, INC XTACK-160-H AF04669

Patients

Seq Age Sex Outcome Treatment
1 38 YR Male Required Intervention