INSYTE AUTOG BC GLOBAL BL 22GA X 1.0IN
Report
- Report Number
- 1710034-2023-01412
- Event Type
- Malfunction
- Date Received
- December 7, 2023
- Date of Event
- November 14, 2023
- Report Date
- March 25, 2024
- Manufacturer
- BECTON DICKINSON
- Product Code
- ONB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 381023 AND LOT NUMBER 3240090. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.
DEVICE PROBLEM CODE: A0510 - RETRACTION PROBLEM PATIENT PROBLEM CODE: F26 ¿ NO HEALTH CONSEQUENCES OR IMPACT PR (B)(4): INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED.
MATERIAL :381023 UPDATED AND BATCH:3240090.
MATERIAL#: 644696, BATCH#: UNKNOWN(3240090) PROVIDED AS UNKNOWN. IT WAS REPORTED BY CUSTOMER THAT SAFETY BUTTON ON IV CATHETER DOES NOT IMMEDIATELY CAUSE RETRACTION OF NEEDLE INTO HANDLE WHEN DEPRESSED. THE NEEDLE WILL RETRACT WITH REPEATED ATTEMPTS AT DEPRESSING THE BUTTON. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. SAFETY BUTTON ON IV CATHETER DOES NOT IMMEDIATELY CAUSE RETRACTION OF NEEDLE INTO HANDLE WHEN DEPRESSED. THE NEEDLE WILL RETRACT WITH REPEATED ATTEMPTS AT DEPRESSING THE BUTTON. ITEM#644696. LOT#3240090.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 213011 | INSYTE AUTOG BC GLOBAL BL 22GA X 1.0IN | CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS | ONB | BECTON DICKINSON | 3240090 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |