TRUE METRIX AIR
Report
- Report Number
- 1000113657-2023-00590
- Event Type
- Malfunction
- Date Received
- December 7, 2023
- Date of Event
- October 27, 2023
- Report Date
- December 7, 2023
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE RETURNED FOR EVALUATION. PRODUCT TESTING WAS PERFORMED AND DEFECT FOUND ON RETURNED TEST STRIPS: HIGH RESULTS. NO DEFECT FOUND ON RETURNED METER. ROOT CAUSE: RC-072: VIAL LEFT OPEN FOR AN EXTENDED PERIOD OF TIME. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 09-NOV-2023 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 143, 124, 118 AND 147 MG/DL. THE CUSTOMER'S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT IS 114 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULT OF 123 MG/DL USING TRUE METRIX AIR METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 11/17/2023 AND TESTS STRIPS WERE OPENED ONE MONTH PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE NOT SET): RESULT 1: 143 MG/DL , DATE: (B)(6) 2023 , TIME: 3:52 AM FASTING. RESULT 2: 124 MG/DL DATE: (B)(6) 2023, TIME: 3:50 AM FASTING. RESULT 3: 118 MG/DL,, DATE: (B)(6) 2023, TIME: 4:23 AM FASTING RESULT 4: 118 MG/DL , DATE: (B)(6) 2023, TIME: 4:22 AM FASTING. RESULT 5: 147 MG/DL , DATE: (B)(6) 2023, TIME: 4:20 AM FASTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1127566 | TRUE METRIX AIR | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TRUE METRIX NFRS24/CS 50CT MG/DL | ZA4777S | 00021292007836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |