ANTIROTATION SCREW FOR FEMORAL NECK SYS 85MM LENGTH - STERIL
Report
- Report Number
- 8030965-2023-15424
- Event Type
- Injury
- Date Received
- December 6, 2023
- Date of Event
- November 14, 2023
- Manufacturer
- SYNTHES GMBH
- Product Code
- KTT
- UDI-DI
- 07612334089953
- PMA / PMN Number
- K172872
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D9: COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. E1: INITIAL REPORTER FACILITY NAME: (B)(6) MEDICAL CENTER FOUNDATION E3: REPORTER IS A J&J EMPLOYEE. H3, H4, H6: PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE DEVICE HISTORY REVIEW (DHR): A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE#:04.168.485S. LOT #:3189P20. IT WAS ELECTRONICALLY REVIEWED AND NO NONCONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS. THE PRODUCT WAS RELEASED ON: 28/11/2022. MANUFACTURING SITE: JABIL GRENCHEN. EXPIRY DATE: 01/11/2032. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT OPEN REDUCTION/INTERNAL FIXATION SURGERY WITH THE FNS (FEMORAL NECK SYSTEM) FOR A RIGHT FEMORAL NECK FRACTURE. ON (B)(6) 2023, THE PATIENT FELL, AND AN X-RAY SHOWED THAT CUTOUT OCCURRED. THEREFORE, A THA (TOTAL HIP ARTHROPLASTY) REVISION SURGERY WAS PLANNED TO BE PERFORMED ON (B)(6) 2023. THIS REPORT INVOLVES ONE ANTIROTATION SCREW FOR FEMORAL NECK SYS 85MM LENGTH - STERIL. THIS IS REPORT 3 OF 4 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2179848 | ANTIROTATION SCREW FOR FEMORAL NECK SYS 85MM LENGTH - STERIL | APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT | KTT | SYNTHES GMBH | 3189P20 | 07612334089953 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Female | Required Intervention | BOLT F/FEM NECK SYST F/CONSTRUCT LENGTH| LOCKSCR Ø5 SELF-TAP L38 TAN| PL 1-HO F/FEM NECK SYST TAN |