FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL

MDR report key: 1827296 · Received September 7, 2010

Report

Report Number
6000001-2010-03055
Event Type
Malfunction
Date Received
September 7, 2010
Date of Event
July 23, 2010
Report Date
July 27, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THERE IS AN ONGOING CAPA INVESTIGATION, MDQ-CAPA-(B)(4) ASSOCIATED WITH THIS REPORT. (B)(4)

Additional Manufacturer Narrative · 1

THE REPORTED CONDITION OF FAILURE CODE 808:02 WAS CONFIRMED BUT NOT DUPLICATED. THE REPORTED CONDITION IS DUE TO A FAULTY PHM (PUMP HEAD MODULE). THIS DEVICE WILL NOT BE RETURNING TO THE CUSTOMER , THEREFORE NO REPAIRS WILL BE MADE. (B)(4)

Description of Event or Problem · 1

THE CUSTOMER REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 808:02. THE REPORTED CONDITION WAS IDENTIFIED DURING BIOMEDICAL TESTING. THERE WAS NO DOCUMENTED PATIENT INJURY OR MEDICAL INTERVENTION RELATED TO THE REPORTED CONDITION. THE PUMP WAS CATEGORIZED AS A COLLEAGUE 2006 PUMP WITH SOFTWARE VERSION 5.09.90. NO ADDITIONAL INFORMATION IS AVAILABLE. DURING BAXTER QUALITY ENGINEERING REVIEW OF THE EVENT HISTORY ON (B)(4), 2010, IT WAS DETERMINED THAT THE REPORTED CONDITION OCCURRED DURING DELIVERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1