NEEDLE SFTYGLD 25X1 RB MCK
Report
- Report Number
- 1213809-2023-01402
- Event Type
- Malfunction
- Date Received
- December 5, 2023
- Date of Event
- November 13, 2023
- Report Date
- February 21, 2024
- Manufacturer
- BECTON DICKINSON
- Product Code
- FMI
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PR 9260311 ¿ FOLLOW UP MDR FOR DEVICE EVALUATION. DURING REVIEW OF THE PROCESSES, IT WAS DETERMINED PROCESS VARIATIONS DURING THE NEEDLE LUBRICANT APPLICATION CAN CREATE CLOGGED NEEDLES. IT COULD BE POSSIBLE SOME SAMPLES ARE ESCAPES FROM THE INCIDENT THAT OCCURRED DURING PRODUCTION. CORRECTIVE AND PREVENTATIVE ACTIONS HAVE BEEN IMPLEMENTED. SEVERAL QUALITY INITIATIVES HAVE BEEN IMPLEMENTED ON OUR MANUFACTURING LINE TO ENSURE THAT THE NEEDLE LUBRICANT APPLICATION IS PROPERLY APPLIED DURING THE MANUFACTURING PROCESS. ADDITIONALLY, A MONITORING PROGRAM IS ALSO IN PLACE TO VERIFY THE NEEDLE LUBRICANT IS APPLIED UNIFORMLY TO THE NEEDLE. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
DEVICE PROBLEM CODE: A140901 - COMPLETE BLOCKAGE. PATIENT PROBLEM CODE: F26 ¿ NO HEALTH CONSEQUENCES OR IMPACT PR (B)(4): INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED. THE ACTUAL DATE OF EVENT IS UNKNOWN. THE DATE RECEIVED BY MANUFACTURER WAS ENTERED INTO THE DATE OF EVENT FIELD.
NO ADDITIONAL INFORMATION.
MATERIAL#: UNKNOWN BATCH#: 2202913. IT WAS REPORTED BY CUSTOMER THAT THE PROBLEM IS THAT WE CANNOT INJECT VACCINE THROUGH THE NEEDLE. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. THE PROBLEM IS THAT WE CANNOT INJECT VACCINE THROUGH THE NEEDLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2049442 | NEEDLE SFTYGLD 25X1 RB MCK | NEEDLE, HYPODERMIC, SINGLE LUMEN | FMI | BECTON DICKINSON | 2202913 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |