FDA Adverse Event Death Summary report: N

HIPOINT 88 ACCESS SYSTEM (088 ACCESS SYSTEM)

MDR report key: 18269049 · Received December 5, 2023

Report

Report Number
3016522967-2023-00005
Event Type
Death
Date Received
December 5, 2023
Date of Event
November 9, 2023
Report Date
November 26, 2023
Manufacturer
ROUTE 92 MEDICAL, INC.
Product Code
QJP
PMA / PMN Number
K201518
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Description of Event or Problem · 0

PATIENT EXPERIENCING STROKE DUE TO DISTAL OCCLUSION OF M1 SEGMENT OF THE MCA. MODERATE VESSEL TORTUOSITY. PHYSICIAN ADVANCED HIPOINT 88 WITH TENZING 8 THROUGH BASE CAMP SHEATH SYSTEM TO THE MID M1 WITH SOME DIFFICULTY. M1 VESSEL WAS SHAPED LIKE A HORSESHOE. TENZING 8 WENT EASILY TO THE DISTAL OCCLUSION, BUT THE HIPOINT 88 TIP WOULD NOT ADVANCE PAST THE MID M1 SEGMENT. THE PHYSICIAN ASPIRATED FOR 2 MINUTES EVEN THOUGH THE TIP WAS NOT AT THE CLOT. A SUBSEQUENT ANGIOGRAM INDICATED A VESSEL PERFORATION IN THE PROXIMAL SEGMENT OF THE M1. VESSEL SEALED AFTER BALLOON TAMPONADE OF VESSEL. THROMBECTOMY PROCEDURE ABORTED AND PATIENT TRANSFERRED TO ICA FOR FURTHER CARE/OBSERVATION. IT WAS LATER COMMUNICATED THAT THE PATIENT EXPIRED. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2235508 HIPOINT 88 ACCESS SYSTEM (088 ACCESS SYSTEM) PERCUTANEOUS CATHETER QJP ROUTE 92 MEDICAL, INC. 23101902

Patients

Seq Age Sex Outcome Treatment
1 72 YR Male Death BASE CAMP SHEATH SYSTEM| HIPOINT 88 SYSTEM