STEM: QUADRA-P QUADRA-P LAT STEM SIZE 7
Report
- Report Number
- 3005180920-2023-00979
- Event Type
- Injury
- Date Received
- December 5, 2023
- Date of Event
- August 22, 2023
- Report Date
- January 30, 2024
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630040720595
- PMA / PMN Number
- K181254
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- 003
Narratives
CLINICAL EVALUATION PERFORMED BY CLINICAL AFFAIRS DEPARTMENT ON (B)(6) 2024 ALMOST 1 YEAR AFTER A PRIMARY THA, USING QUADRA-P IMPLANT, STEM RADIOLUCENCIES WERE NOTICED BY THE SURGEON DURING A POST-OP X-RAY CHECK. NO CLINICAL INFORMATION ABOUT THE PATIENT WAS REPORTED AND NO REVISION SURGERY IS PLANNED. FROM THE AVAILABLE X-RAYS IMAGES, RADIOLUCENCY LINES ARE APPRECIABLE AROUND THE PROXIMAL FEMUR, IN BOTH THE AP AND LATERAL VIEWS, AND ALSO A SLIGHTLY SHORTENED LEG (OR PERHAPS SUBSIDENCE OF THE STEM SECONDARY TO LOOSENING). AS THE IMMEDIATE POST-OP IMAGES ARE UNAVAILABLE, NO COMMENT CAN BE MADE ON THE DEVELOPMENT OF THE RADIOLUCENT LINES. THIS SHORT TERM OUTCOME IS VERY RARE AND SURPRISING AND WE ARE UNABLE TO MAKE A REASONABLE FORECAST OF THE POSSIBLE EVOLUTION.
BATCH REVIEW PERFORMED ON 16 NOVEMBER 2023. LOT 2000333: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 20-MAY-2020. EXPIRATION DATE: 2025-MAY-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH 1 SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.
THE SURGEON NOTED RADIOLUCENT LINES ON RADIOGRAPHS 1 YEAR AFTER SURGERY AND DIAGNOSED MILD STEM LOOSENING ON THE LEFT HIP IN ZONE 1/7/8. THE PATIENT DOES NOT COMPLAIN OF ANY PAIN OR DISCOMFORT. REVISION SURGERY IS NOT PLANNED. THE PATIENT WILL BE MONITORED. NOTE: PATIENT IS BILATERAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2204921 | STEM: QUADRA-P QUADRA-P LAT STEM SIZE 7 | CEMENTLESS STEM | LZO | MEDACTA INTERNATIONAL SA | 2000333 | 07630040720595 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Male | Required Intervention |