FDA Adverse Event
Malfunction
Summary report: N
NURSE ASSIST, LLC
MDR report key: 18255608
·
Received December 4, 2023
Report
- Report Number
- 3002695476-2023-00008
- Event Type
- Malfunction
- Date Received
- December 4, 2023
- Date of Event
- April 28, 2023
- Report Date
- December 4, 2023
- Manufacturer
- NURSE ASSIST, LLC
- Product Code
- FRO
- UDI-DI
- 40884389338248
- PMA / PMN Number
- K083042
- Removal / Correction Number
- 1650927-3OCT2023-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE EXTENDED TIME BETWEEN DATE OF EVENT AND DATE OF THIS REPORT IS DUE TO NURSE ASSIST RECENTLY RECLASSIFYING THIS TYPE OF OCCURRENCE AS HAVING THE POTENTIAL FOR CAUSING OR CONTRIBUTING TO A DEATH OR A SERIOUS INJURY IF THE MALFUNCTION WERE TO RECUR. NURSE ASSIST IS CONDUCTING A RETROSPECTIVE REVIEW OF PRODUCT COMPLAINTS TO IDENTIFY ALL OCCURRENCES OF MALFUNCTION OF THE STERILE BARRIER.
Description of Event or Problem · 0
ON 28-APR-2022, CUSTOMER NOTIFIED NURSE ASSIST VIA E-MAIL OF THE FOLLOWING EVENT DESCRIPTION FROM A CUSTOMER: RECEIVED BOTTLES OF RDI30296 THAT WERE LEAKING. THERE WAS NO SHIPPING DAMAGE. THESE BOTTLES WERE REPORTED BY THE INITIAL REPORTER AS LEAKING PRIOR TO USE ON A PATIENT. THERE WAS NO REPORT OF ANY PATIENT IMPACT OR ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1451698 | NURSE ASSIST, LLC | 0.9% NORMAL SALINE FOR IRRIGATION USP, IN SCREW TOP BOTTLE | FRO | NURSE ASSIST, LLC | RDI30296 | 22073659 | 40884389338248 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |