Report Number
3002682307-2023-00347
Event Type
Malfunction
Date Received
December 2, 2023
Date of Event
May 12, 2023
Report Date
January 10, 2024
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
UDI-DI
30382903027720
PMA / PMN Number
UKN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302772 AND LOT NUMBER 2110403. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR RETURN, TWENTY (20) RETAINED SAMPLES WERE OBTAINED FROM THE MANUFACTURING FACILITY. THE RETAINED SAMPLES WERE VISUALLY EXAMINED AND NO SIGNS OF DEFECT WERE IDENTIFIED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE COULD NOT BE DETERMINED FOR THIS INCIDENT. AT THIS TIME, FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY. OUR QUALITY TEAM WILL CONTINUE TO CLOSELY MONITOR THE MANUFACTURING PROCESS FOR SIGNS OF THIS POTENTIAL DEFECT AND ANY EMERGING TRENDS.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD SYRINGE SOLOMED HAD PACKAGING PROBLEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: BROKEN SYRINGE PACKAGING WAS FOUND BEFORE THE CHILD WAS ADMINISTERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1504268 BD SYRINGE SOLOMED SYRINGE, PISTON FMF BECTON DICKINSON, S.A. 2110403 30382903027720