FDA Adverse Event Malfunction Summary report: N

CARDIOQUIP MODULAR COOLER HEATER

MDR report key: 18252058 · Received December 1, 2023

Report

Report Number
3007899424-2023-00468
Event Type
Malfunction
Date Received
December 1, 2023
Date of Event
October 30, 2023
Report Date
July 26, 2024
Manufacturer
CARDIOQUIP LLC
Product Code
DWC
PMA / PMN Number
K102147
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

UDI RELATED DATA QUALITY UPDATES ONLY.

Additional Manufacturer Narrative · 0

FOLLOWING TEST RESULTS OUTSIDE OF CARDIOQUIP SPECIFICATIONS, CARDIOQUIP RECOMMENDED THAT THE DEVICE RECEIVE AN INTERNAL WATER PATHWAY REPLACEMENT. THE CUSTOMER THEN SENT THE DEVICE TO CARDIOQUIP FOR REPAIR. THE DEVICE WAS RETURNED TO SPECIFICATION VIA AN INTERNAL WATER PATHWAY REPLACEMENT. FOLLOWING THE REPAIR, THE DEVICE PASSED INSPECTION AND IS FULLY FUNCTIONAL.

Additional Manufacturer Narrative · 0

A TESTS OF WATER QUALITY RESULTS WAS FOUND TO BE OUTSIDE OF CARDIOQUIP'S ACCEPTABLE LIMITS. THIS INFORMATION WAS RELAYED TO THE CUSTOMER VIA EMAIL ON 11/02/2023. ALONG WITH THIS THE CUSTOMER WAS INFORMED THAT THEY COULD EITHER COMPLETE AN INTERNAL WATER PATHWAY REPLACEMENT ON THEIR DEVICE TO RETURN IT TO SPECIFICATION OR PARTICIPATE IN CARDIOQUIP'S TRADE-IN PROGRAM. AS OF THE DATE OF THIS REPORT, THERE HAS BEEN NO RESPONSE TO THESE RECOMMENDATIONS. A FOLLOW-UP WILL BE FILED IF ANY ADDITIONAL INFORMATION OR INFORMATION THAT CORRECTS THIS REPORT IS OBTAINED.

Description of Event or Problem · 0

LAB RESULTS (ON (B)(6) 2023): (B)(6) 738,000 MPN/ML (ABOVE ACCEPTABLE LIMIT). TO REMEDIATE THE DEVICE CONTAMINATION, THE CUSTOMER EITHER PROCEED WITH AN INTERNAL WATER PATH REPLACEMENT PROCEDURE OR A TRADE-IN PROGRAM.

Description of Event or Problem · 0

LAB RESULTS (B)(6) 2023): 10161042 - 738,000 MPN/ML (ABOVE ACCEPTABLE LIMIT). TO REMEDIATE THE DEVICE CONTAMINATION, THE CUSTOMER EITHER PROCEED WITH AN INTERNAL WATER PATH REPLACEMENT PROCEDURE OR A TRADE-IN PROGRAM.

Description of Event or Problem · 0

LAB RESULTS ((B)(6) 2023): (B)(4) - 738,000 MPN/ML (ABOVE ACCEPTABLE LIMIT). TO REMEDIATE THE DEVICE CONTAMINATION, THE CUSTOMER EITHER PROCEED WITH AN INTERNAL WATER PATH REPLACEMENT PROCEDURE OR A TRADE-IN PROGRAM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1505168 CARDIOQUIP MODULAR COOLER HEATER CARDIOPULMONARY BYPASS DEVICE DWC CARDIOQUIP LLC MCH-1000(I)

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown