FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 18251838 · Received December 1, 2023

Report

Report Number
2647876-2023-00613
Event Type
Malfunction
Date Received
December 1, 2023
Date of Event
October 30, 2023
Report Date
December 8, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420239
PMA / PMN Number
K113558
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

G5: PMA/510(K)#: K222591 H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: CATALOG: 442023. BATCH NO.: 3130646. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS.

Description of Event or Problem · 0

REPORT 2 OF 2. IT WAS REPORTED THAT WHEN USING THE BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC), THERE WERE TWO MOLECULAR FALSE POSITIVES. NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " REPORTED 2 MOLECULAR FP WHICH WAS C.TROPICALIS WITH BCID 2. GRAIN STAIN WAS GPB AND CULTURE WAS CORYNEBATERIUM. (2) GVB & GPCCL, CULTURE S CAPITIS & PAENIBACILLUS."

Description of Event or Problem · 0

REPORT 2 OF 2. IT WAS REPORTED THAT WHEN USING THE BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC), THERE WERE TWO MOLECULAR FALSE POSITIVES. NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " REPORTED 2 MOLECULAR FP WHICH WAS C.TROPICALIS WITH BCID 2. GRAIN STAIN WAS GPB AND CULTURE WAS CORYNEBATERIUM. (2) GVB & GPCCL, CULTURE S CAPITIS & PAENIBACILLUS."