FDA Adverse Event Malfunction Summary report: N

CTR73, 12X100 KII OPT ZTHR 6/BX

MDR report key: 18250378 · Received December 1, 2023

Report

Report Number
2027111-2023-00790
Event Type
Malfunction
Date Received
December 1, 2023
Date of Event
October 6, 2023
Report Date
March 20, 2024
Manufacturer
APPLIED MEDICAL RESOURCES
Product Code
GCJ
UDI-DI
00607915123680
PMA / PMN Number
K060096
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE EVENT DEVICE IS ANTICIPATED TO RETURN. A FOLLOW-UP REPORT WILL BE PROVIDED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 0

THE EVENT UNIT WAS RETURNED TO APPLIED MEDICAL FOR EVALUATION. VISUAL INSPECTION CONFIRMED THE PRESENCE OF HAIR INSIDE THE STERILE UNOPENED POUCH. BASED ON THE CONDITION OF THE RETURNED UNIT AND THE DESCRIPTION OF THE EVENT, THE HAIR LIKELY ORIGINATED FROM AN OPERATOR DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 0

NAME OF PROCEDURE BEING PERFORMED: NA (INCOMING INSPECTION). DETAILED DESCRIPTION OF EVENT: [DISTRIBUTOR] INCOMING INSPECTION PO #(B)(4). THIS IS A COMPLAINT FROM [DISTRIBUTOR] EVALUATION TEAM. THE LOT MENTIONED ABOVE WAS FOUND TO HAVE DEFECTIVE UNITS PER OUR INCOMING INSPECTION. DELIVERY DATE: OCTOBER 6, 2023. WE FOUND DEFECTS AS THE RESULT OF INCOMING INSPECTION OF THE PRODUCTS. PLEASE REFER TO THE ATTACHMENT MATERIAL FOR THE QUANTITY WHICH RETURNS THE PRODUCTS. ) WE FOUND THE DEFECTS THAT ALREADY INFORMED APPLIED MEDICAL BY CER. WE JUST INFORM YOU THE DEFECT AND RETURN THE PRODUCTS FOR FOLLOWING DEFECTS. YOU DO NOT NEED INVESTIGATION. HAIRLIKE OBJECT IN POUCH - CTR73 LOT #1495004 (2 EA.). TYPE OF INTERVENTION: NA (INCOMING INSPECTION). PATIENT STATUS: NA (INCOMING INSPECTION).

Description of Event or Problem · 0

NAME OF PROCEDURE BEING PERFORMED NA (INCOMING INSPECTION). DETAILED DESCRIPTION OF EVENT #2023-003419 - HAIRLIKE OBJECT IN POUCH - CTR73 LOT #1495004 ((B)(4) EA). [DISTRIBUTOR] INCOMING INSPECTION PO # (B)(4). THIS IS A COMPLAINT FROM [DISTRIBUTOR] EVALUATION TEAM. THE LOT MENTIONED ABOVE WAS FOUND TO HAVE DEFECTIVE UNITS PER OUR INCOMING INSPECTION. DELIVERY DATE: OCTOBER 6,2023 PLEASE REFER TO THE FILE. WE FOUND DEFECTS AS THE RESULT OF INCOMING INSPECTION OF THE PRODUCTS. PLEASE REFER TO THE ATTACHMENT MATERIAL FOR THE QUANTITY WHICH RETURNS THE PRODUCTS. WE FOUND THE DEFECTS THAT ALREADY INFORMED APPLIED MEDICAL BY CER. WE JUST INFORM YOU THE DEFECT AND RETURN THE PRODUCTS FOR FOLLOWING DEFECTS. YOU DO NOT NEED INVESTIGATION. HAIRLIKE OBJECT IN POUCH - CTR73 LOT #1495004 ((B)(4) EA) PATIENT STATUS: NA (NO PATIENT INVOLVEMENT). TYPE OF INTERVENTION: NA (INCOMING INSPECTION).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1451425 CTR73, 12X100 KII OPT ZTHR 6/BX LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ APPLIED MEDICAL RESOURCES CTR73 1495004 00607915123680

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown