FDA Adverse Event Injury Summary report: N

REVERSE SHOULDER SYSTEM 04.01.0122 HUMERAL REVERSE HC LINER Ø39/+0MM

MDR report key: 18245853 · Received December 1, 2023

Report

Report Number
3005180920-2023-00933
Event Type
Injury
Date Received
December 1, 2023
Date of Event
October 31, 2023
Report Date
December 1, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
PHX
UDI-DI
07630040706292
PMA / PMN Number
K170452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 09 NOVEMEBR 2023: LOT 2008853: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18-MAR-2021. EXPIRATION DATE: 2025-11-05. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL COMPONENT INVOLVED (NOT REVISED): REVERSE SHOULDER SYSTEM 04.01.0110 HUMERAL REVERSE METAPHYSIS +0MM/0° (K170452) LOT 2114012: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-FEB-2022. EXPIRATION DATE: 2027-01-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. REVERSE SHOULDER SYSTEM 04.01.0151 GLENOID BASEPLATE Ø24.5X15 (K170452) LOT 2116432: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 08-MAR-2022. EXPIRATION DATE: 2027-02-22. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. REVERSE SHOULDER SYSTEM 04.01.0162 GLENOID POLYAXIAL LOCKING SCREW - L34 (K170452) LOT 2010019: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 27-NOV-2020. EXPIRATION DATE: 2025-11-16. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. REVERSE SHOULDER SYSTEM 04.01.0162 GLENOID POLYAXIAL LOCKING SCREW - L34 (K170452) LOT 2118708: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 07-FEB-2022. EXPIRATION DATE: 2027-01-24. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. REVERSE SHOULDER SYSTEM 04.01.0170 GLENOSPHERE 39XØ24.5 LOT 2003810B (K170452): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11-OCT-2021. EXPIRATION DATE: 2026-09-27. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

THE PATIENT HAD A PRIMARY SHOULDER SURGERY ON (B)(6) 2022. ON (B)(6) 2022, THE PATIENT CAME IN REPORTING PAIN DUE TO FALLING AND DISLOCATING THE LINER FROM THE GLENOSPHERE. THE SURGEON REVISED THE LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER, METAPHYSIS, BASEPLATE, GLENOSPHERE AND SCREWS. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1633062 REVERSE SHOULDER SYSTEM 04.01.0122 HUMERAL REVERSE HC LINER Ø39/+0MM REVERSE HUMERAL LINER PHX MEDACTA INTERNATIONAL SA 2008853 07630040706292

Patients

Seq Age Sex Outcome Treatment
1 71 YR Male Required Intervention