FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1823916 · Received January 16, 2008

Report

Report Number
1527736-2008-00310
Event Type
Malfunction
Date Received
January 16, 2008
Date of Event
January 1, 2007
Report Date
December 19, 2007
Manufacturer
ETHICON ENDO-SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVAL SUMMARY: THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT. IT WAS TESTED ON A GENERATOR AND NO ERROR CODE WAS NOTED. HOWEVER, A HISSING SOUND WAS HEARD. THE HAND PIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. ADDITIONALLY, THE INSTRUMENT NO LONGER MEETS THE SPECIFICATION FOR CONTINUITY. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THE HANDPIECE DOES NOT WORK. NO OTHER INFO AVAILABLE. NO PT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE LFL ETHICON ENDO-SURGERY, INC. (CINCINNATI) NA NA

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR| HANDPIECE